PAGE 14
DIAGENODE
DOUBLE HEAD PERISTALTIC PUMP
USER MANUAL
Europe Diagenode sa
/ CHU - Tour GIGA - B34 - 3e étage // Avenue de l’Hôpital, 1 // 4000 Liège (Sart Tilman) // Belgium // Phone: (+32) 4 364 20 50 // Mail: info
@
diagenode.com
Instructions for use
Identification of Operating Controls
Procedure for Setting
Start the cooling system; wait until the water is at expected temperature.
Consult cooling system’s manual if necessary.
Rough water interferes in sonication phenomenon. Bioruptor’s sonication bath does not allow water movement upper
than 500 ml/min. Income water flow (cold flow) should be adjusted around 450 ml/ml. The back flow (warm flow) must
be at least the same of the income flow.
Note:
The drawing flow cannot be lower than the discharging flow otherwise the volume of water in bath would increase upper
to the limit. On the other hand, the warm flow is able to be higher than the cold flow because once the water upper the
level is drawn, pumps will draw air which is not a problem.
Once the flows adjusted, you don’t need to change them, just switch off the main switch.
Pump Control graduated
from 0 to 600 ml/min
Power Inlet
Main Switch
Signalitic board