DOUBLE HEAD PERISTALTIC PUMP
USER MANUAL
Europe Diagenode sa
/ CHU - Tour GIGA - B34 - 3e étage // Avenue de l’Hôpital, 1 // 4000 Liège (Sart Tilman) // Belgium // Phone: (+32) 4 364 20 50 // Mail: info
@
diagenode.com
Introduction
Warnings
These instructions cannot claim to cover all details of possible equipment variations, nor in particular can they provide
for every possible example of installation or operation.
Trouble-free and safe operation of the unit is dependent on proper transport and installation by qualified personnel.
Guarantee
Limited one year global warranty. Diagenode guarantees its products against possible manufacturing defects in material
and workmanship. Diagenode products are rigorously tested to ensure that the products you trust meet stringent
standards. Consequently, if a problem occurs with a Diagenode product and the problem is caused by manufacturing
defects in material and workmanship, Diagenode will, in its discretion, either fix or replace the product in accordance
with the warranty terms and conditions stated herein. The warranty applies only to the first purchaser of the product for
a period of one year starting from the date it was delivered.
In case of repair or replacement on a product under warranty, expenses will be at Diagenode’s charge, including any
costs required to return the repaired or replacement product to you. This warranty covers only manufacturing defects
and does not cover any damage caused by misuse (non-respect of recommendations described in this manual), neglect,
accidents, abrasion, exposure to extreme temperatures, solvents, acids.
We strongly recommend that maintenances or repairs on your product are performed by our approved Diagenode service
center. Improper or incorrectly performed maintenance or repairs void this warranty.
Intended use of the equipment
The pump device as well as all components from this parcel, including plastic nozzle connectors, tubing, strips and
tube holders can be used as a junction between any Water Reservoir Chiller (min specification mentioned below) and a
Bioruptor waterbath (Bioruptor®: one waterbath, Bioruptor® Twin: two waterbaths).
Any brand – any type: Bath circulator, immersion cooler,… – of Water Reservoir Chiller is suitable as long as:
- the capacity of the reservoir is 4 liters (min)
- the reservoir allows insertion of hoses
- the cooling power is 500 Watt (min) at external temperature of 20°C
- the cooling temperature range reaches 4°C (+/- 3°C)
Warning:
Recirculating Chiller device (close loop without possibility of water reservoir) is not usable.
This equipment can be used to connect two Sonication Baths to a Water Reservoir Chiller.
This equipment is designed to be used for the above described application only. Any medical application is out of scope.