![background image](http://html.mh-extra.com/html/diagenode/b02030001-twi-pump/b02030001-twi-pump_user-manual_2492808006.webp)
PAGE 6
DIAGENODE
DOUBLE HEAD PERISTALTIC PUMP
USER MANUAL
Europe Diagenode sa
/ CHU - Tour GIGA - B34 - 3e étage // Avenue de l’Hôpital, 1 // 4000 Liège (Sart Tilman) // Belgium // Phone: (+32) 4 364 20 50 // Mail: info
@
diagenode.com
Technical Specifications
Item
Description
Dimensions
100 x 200 x 250 mm
Weight
2.128 kg
Motor type
DC Geared Motor 8 - 4 VDC
RPM
0 - 210 round per minute
Discharge rate
0 - 610 ml per minute
Discharge pressure
0,1 MegaPascal
Suction Height
1,5 meter
Tube size (inch)
1/4 (Inner Diameter) x 3/8 (Outer Diameter)
Tube size (mm)
6,35 (Inner Diameter) x 9,525 (Outer Diameter)
Tube matter
Pharmed - Silicone - Fluran - Tygon LFL
Power supply single phase
Input
Rated voltage Vin
100-240 V AC
Voltage range
85…264 V AC
Line frequency range
47…63 Hz
Rated current Iin
0,7 - 0,35 A
Double fused
1 A
Internal circuits
Rated voltage V
24 V DC
Rated current
1,3 A
Current limitation
2 A
Note:
In case of infection, the entire water circulation loop (tubing) can be sterilised by autoclave or ethylene oxide as well as
by gamma radiation.
Environmental Conditions
Devices are designed to be safe under the following conditions:
• indoor use;
• altitude up to 2 000 m;
• operating external temperature 0°C to 40°C;
• fluid temperature 4°C to 50°C
• maximum relative humidity 80% for temperatures up to 31°C decreasing linearly to 50% relative humidity at 40°C,
without condensation.
• MAINS supply voltage from 100V AC to 230V AC with fluctuations up to ±10% of the nominal voltage;
• transient overvoltages typically present on the MAINS supply;
• Degree of protection: IP20;
• POLLUTION DEGREE 2 (Normally only non-conductive POLLUTION occurs. Occasionally, however, a temporary
conductivity caused by condensation is expected)
Warning:
Never use in atmosphere with flammable gas. Pump motors may generate sparkles and may light fire or explosion in
atmosphere with flammable gas.
Never use in any location where there is a possibility of extreme dust.
Environment exempt from sunlight is required.