Service/maintenance
OPERATION MANUAL
Chapter 7
Water Cooler
Errors and omissions excepted.
V1.1.0en/13.12.12
44
7.1
Display error messages
In case of malfunction alarms and warning messages are indicated through the display.
Display
Cause
Effect, measurement
Flashing display of
temperature value
Warning: Over- or un-
dertemperature.
Thermoregulation continues.
Limit value from set point:
Minichiller (Plus)
− 5 K/+ 2 K
Minichiller w (Plus) − 5 K/+ 2K
Minichiller H (Plus) ± 5 K
F1 – flashing
Malefunction sensor F1
Broken or short circuit
Thermoregulation is inactiv.
(Pump off, compressor off, heating off)
Check the Sensor.
E1 – flashing
Inlet E1 indicates failure.
No release-signal, level-
alarm.
Thermoregulation is inactiv.
(Pump off, compressor off, heating off)
Check level.
Reset only possible if level ok.
E2 – flashing
Inlet E1 indicates failure.
Pump is running and
circulation is missing or
pump is running and
water pressure is missing.
Thermoregulation is inactiv.
(Pump off, compressor off, heating off)
Reset only possible if disconnected from the power
supply.
E3 – flashing
Even though thermoregu-
lation is off, inlet E1 is
indicating circulation.
Thermoregulation is inactiv.
(Pump off, compressor off, heating off)
Failure is being reset if inlet E1 is open again in
standby.
EP – flashing
Loss of data in parameter
memory.
Thermoregulation is inactiv.
(Pump off, compressor off, heating off)
While the error message is displayed, the display changes constantly between the error message
and the actual value.
Overview of the
menu items
INFORMATION
.
Summary of Contents for BioAcc-cool
Page 1: ...Cat No BioAcc cool Water Cooler Version 1 1 ...
Page 6: ...OPERATION MANUAL Water Cooler Errors and omissions excepted V1 1 0en 13 12 12 6 ...
Page 10: ...OPERATION MANUAL Water Cooler Errors and omissions excepted V1 1 0en 13 12 12 10 ...
Page 58: ......
Page 59: ...PAGE 55 www diagenode com ...
Page 60: ...orders diagenode com orders na diagenode com www diagenode com ...