Typ
Digital thermometer
Range
32,0°C - 42,9°C
Accuracy
±0,1°C, 35,5°C - 42,0°C
±0,2°C - 35,5°C and over 42,0°C
at ambient temperature of 18°C - 28°C
Display
LCD 3 i 1/2 digits
Memory
Storing the last measured value
Battery
One 1,5 V DC (LR41 or SR41, UCC 392)
Battery life
approx. 200 hours
Dimension
12,4cm x 1,8cm x 0,9cm (L x W x H)
Weight
Approx. 10 g incuding battery
Operating environment
5°C~40°C, 15%~95% R.H.
Storage and Transportation -20°C ~ 55°C, 15% ~ 95%, R.H.
Waterpoof level
IP 22
Clasification
Typ BF
Thank you for purchasing the Diagnostic T-01 thermometer.
The device is designed to be used for measuring the body
temperature of children and adults. Read the instruction manual
carefully before using this device.
Read instructions thoroughly before using digital thermometer.
Choking Hazard: Thermometer cap and battery could be
swallowed. Do not allow children to use this device without
parental supervision.
Do not use thermometer in ear. Designed to be used for oral,
rectal, and armpit (axilla) readings only.
Do not place thermometer battery near extreme heat as it may
explode.
Note: Use of the probe cover may result in a 0.1
°
(0.2℉)
C
discrepancy from actual temperature.
It is recommended, .the performance of the digital
thermometer should accuracy every two years.
Remove battery from the device when not in operation for a
long time.
The use of temperature readings for self-diagnosis is
dangerous. Consult your doctor for the interpretation of results.
Self-diagnosis may lead to the worsening of existing conditions.
Do not attempt measurements when the thermometer is wet
as inaccurate readings may result.
Do not bite the thermometer. Doing so may lead to breakage
and/or injury.
Do not attempt to disassemble or repair the thermometer.
Doing so may result in inaccurate readings.
After each use, disinfect the thermometer especially in case
the device is used by more than one person.
Do not force the thermometer into the rectum. Stop insertion
and the measurement when pain is present. Failure to do so
may lead to injury.
Do not use thermometer orally after being used rectally.
For children who are two years old or younger, please do not
use the devices orally.
DIGITAL THERMOMETER
T-01
WARNING
PLEASE READ CAREFULLY BEFORE USING
This digital thermometer provides a quick and highly accurate reading of an
individual's body temperature. To better understand its functions and to
provide years of dependable results, please read all instructions first.
PRODUCT ILLUSTRATION
CONTENTS
1 Thermometer, 1 Instruction Manual, 1 Storage Case
LCD
Battery Cap
ON/OFF
BUTTON
PROBE
PRECAUTION
Efficiency of the device may
be
reduce
d
if one or more of the
following factors occur:
Operation outside the manufacturer's stated temperature and
humidity range.
Storage outside the manufacturer's stated temperature and
humidity range.
Mechanical shock (e.g. dropping)
Patient temperature is below ambient temperature.
Portable and mobile RF communications can affect the device.
The device needs special pre-cautions regarding EMC according
to the EMC information provided in the accompany documents.
SPECIFICATION
MEASURING
1.
Press the ON/OFF Button
,
a short beep signals the
thermometer ON.
A tone will sound as the screen shows
Fig.1
followed by last recorded temperature. After showing the self-
tested temperature, the thermometer is
now ready for use.
Symbol Lo and
a flashing
°
C
(Fig.2)
indicates the testing mode.
*Note:
If the measured temperature is less than 32
°
C
(90
℉
)
the
LCD will
show
Lo, if the measured temperature is higher than 42.9
°
C
(
109.9℉
)
, the LCD will display Hi.
2. Position thermometer in desired location (mouth, rectum, or
armpit.)
a)
Oral Use: Place
the
thermometer
under tongue
.
Close your mouth
and breathe evenly through the
nose to prevent the measurement
from being influenced by
inhaled/exhaled air. Normal
temperature between 35.7
°
C
-
37.3
°
C
(96.3
℉
- 99.1
℉
)
b)
Rectal Use: Lubricate silver probe
tip with petroleum jelly for easy
insertion. Gently insert
the
sensor
of thermometer
approximately 1cm (less than 1/2") into
rectum. Normal temperature between 36.2
°
C
-
37.7
°
C
(96.2
℉
- 99.9
℉
)
c)
Armpit Use: Wipe armpit
with a
dry
towel
. Place probe in
the
armpit and keep
the
arm pressed firmly at side.
Normal
temperature between 35.2
°
C
-
36.7
°
C (95.4℉
-
98.1℉).
M
c
Fig.1
Fig.2
Fig.3
INSTRUCTION MANUAL
37,1
°
C
37,1
°
C
36,9
°
C
36,9
°
C
36,6
°
C
36,7
°
C
36,5
°
C
EN