O
perator’s manual
version: 2021.06
15
Calibration of the production conductivity electrode
The calibration of the production conductivity electrode is made by taking water directly from the output of
the purified water. The points of calibration must have conductivity values strictly under 10µS/cm.
3.6.8.1
To define a 1st point of calibration
This is the initial case, before any
saving of calibration point.
SELECT:PRESS SET 2 SEC
CALIBRATION
PROD CONDUCTIVITY
MEMBRANE CONDUCTIVITY
SELECT:PRESS SET 2 SEC
CONDUCTIVITY CAL
SET PRODUCTION POINT 1
3.6.8.2
To delete the 1st point of calibration
It is possible to come back to the default conductivity values by deleting the 1
st
point of calibration.
SELECT:PRESS SET 2 SEC
CALIBRATION
PROD CONDUCTIVITY
MEMBRANE CONDUCTIVITY
SELECT:PRESS SET 2 SEC
DELETE PROD. POINT 1
CONDUCTIVITY CAL
SET PRODUCTION POINT 1
SELECT:PRESS SET 2 SEC
CONFIRM DELETION OF
PRODUCTION
POINT 1
Delete the point of
calibration by pressing
with a tool 2 seconds on
the ‘SET’ button
3.6.8.3
To define a 2nd point of calibration
When a 1st point of calibration is
already defined, it is possible to
define a 2nd point.
SELECT:PRESS SET 2 SEC
CALIBRATION
PROD CONDUCTIVITY
MEMBRANE CONDUCTIVITY
SELECT:PRESS SET 2 SEC
CONDUCTIVITY CAL
SET PRODUCTION POINT 1
SET PRODUCTION POINT 2
DELETE PROD. POINT 2
3.6.8.4
To delete the 2nd point of calibration
It is possible to come back to one only point of calibration by deleting the 2
nd
point of calibration.
SELECT:PRESS SET 2 SEC
CALIBRATION
PROD CONDUCTIVITY
MEMBRANE CONDUCTIVITY
SELECT:PRESS SET 2 SEC
SET PRODUCTION POINT 2
CONDUCTIVITY CAL
SET PRODUCTION POINT 1
DELETE PROD. POINT 2
SELECT:PRESS SET 2 SEC
CONFIRM DELETION OF
PRODUCTION
POINT 2
Delete the point of
calibration by pressing
with a tool 2 seconds on
the ‘SET’ button