1.  Transmitting data

Follow the instructions contained in 

the software on your computer for 

transmitting the data. The results are 

transmitted together with their dates and 

times. When you disconnect the cable the 

meter will switch off automatically.

See also: diatesse.nl/support

Creating a data link using Bluetooth® 

(Bluetooth® can be optionally switched 

on or off)

Using Bluetooth®, you can create a data 

link between your meter and a Bluetooth®-

capable smartphone (iOS or Android). 

Download the ProCheck app to your 

smartphone from the App Store (iOS) or from 

Google Play (Android). The ProCheck app 

has been created to assist you simply and 

intuitively in monitoring your blood sugar 

and ketone levels over an extended period. 

For further information, please contact your 

local customer support or distributor. You 

must link your meter and smartphone using 

Bluetooth® before you can transmit data.

Linking your meter and smartphone 

(using a security code)

1.  Turn on Bluetooth® on your smartphone.

2.  Follow the instructions in the ProCheck

  app on your smartphone for connecting 

  to your meter.

3.  Use "Search" to find your meter, which 

  is called DIATESSE XPER xxxx (xxxx = the 

  unique four-digit/letter combination of 

  the Bluetooth® MAC associated with 

  your meter).

4.  Add your meter, press   and press "Yes"

  to save the settings.

5.  Then input your unique code in order to

  complete the connection. Your code is the

  last 7 digits of the meter's serial number,

 

excluding the final digit.

  Input these 6 digits then press "Connect".

6.  Once the devices are linked, the data in

  your meter will be sent via Bluetooth® 

  to the ProCheck app.

Bluetooth® indicator on your meter:

Battery

Your meter comes with two  

1.5 V AAA alkaline batteries.

Low battery warning

The meter will warn you in one of the 

following ways when the power is low.

1.  The symbol "

" appears

 together

  with notifications on the display: the meter

  is working and the results are still accurate,

   but the batteries must be replaced.

2.  The symbol "

" appears together

  with E-b, Error and LOW:

 There is not

  enough power to perform a test. The

  batteries must be replaced immediately.

Replacing the batteries

The meter must be switched off when you 

replace the batteries.

1.  Slide the battery cover to the left to open it.

2.  Remove the 2 empty batteries and insert 

  2 new 1.5 V AAA alkaline batteries.

3.  Close the battery cover.

  If the batteries are correctly inserted, 

  the device will beep.

WARNING:

•  If your meter and computer are

connected by a USB cable then 

your data will be transmitted using 

the cable. You can use Bluetooth® 

to link the devices without using a 

cable, provided that your computer 

supports Bluetooth® and Windows 

10 or higher is loaded.

•  The meter cannot perform a test

  while it is transmitting data.

•  Ensure that your smartphone or

computer supports Bluetooth® Smart 

Technology and that Bluetooth® is 

activated before trying to transmit 

data. Also make sure that your meter 

and the computer or smartphone are 

sufficiently close together so that they 

can link. For the required parameters 

for your (mobile) operating system, 

check the App Store or Google Play 

before downloading and installing 

the app.

•  Bluetooth® functionality is

implemented in various ways by 

different manufacturers. This could 

mean that there are compatibility 

issues between your meter and the 

smartphone.

See, for further information:
diatesse.nl/support

For further information see www.diatesse.nl

WARNING:

You cannot perform tests when the meter  

is connected to a computer.

CONNECTING THE METER 

TO A SMARTPHONE

MAINTENANCE

BLUETOOTH®

INDICATOR

STATUS

Solid blue light

Flashing 

blue light

Bluetooth® is on 

and searching for a 

connection
Bluetooth® linked

Summary of Contents for XPER TD-4289B

Page 1: ...BLOOD GLUCOSE KETONES MONITORING SYSTEM USER MANUAL TD 4289B Version 1 02018 10 DT091018 01F Blood glucose Ketones ...

Page 2: ...o easily measure your own levels wherever and whenever you want Should you have any further queries after reading this manual please contact customer support or your distributor Use This system is intended for exterior use in vitro diagnostic use in a non clinical environment by persons with diabetes and in a clinical environment by healthcare processionals as an aid for assessing the effectivity ...

Page 3: ...tion 7 Meter overview 8 Display 9 Test strip 9 SETTING UP THE METER 10 SETUP MODE 11 For blood sugar testing 11 For β ketone testing 11 QUALITY CONTROLTEST 11 When should you test the device using control solution 11 Performing a test with the control solution 11 TESTING USING BLOOD SAMPLES 12 Preparingthelancingdeviceforabloodtest 12 Preparing the puncture point 12 Performing a blood sugar test 1...

Page 4: ... to receive results that are lower or higher than normal then contact your healthcare professional and follow their advice Only use fresh whole blood samples for testing your blood sugar and β ketone levels Using any other substances will produce incorrect results If you are displaying symptoms that do not match the test results and if you have followed all the instructions in this manual then ple...

Page 5: ... Gen at any time AC before meals PC after meals Automatic quality control mode QC Quality control test using the control solution Average over a number of days Memory symbol Date Time Error warning Alarm symbol Measurement unit Ketone symbol Absorbent opening Test strip grip PLEASE NOTE The front of the test strip must be facing upwards when you insert the test strip The test results could be inco...

Page 6: ...th the control solution when first using the Diatesse XPER When you think that either the meter or the test strips are defective When your test results do not match the way you are feeling or when you believe the results are not accurate When performing a test If you have dropped the meter or have reason to believe it is damaged When the hour minute are flashing press M to select the desired hour ...

Page 7: ...drop must be at least as large as actual size which equals around 0 5 microliter µI volume for a blood sugar test and 0 8 microliter µl volume for a β ketone test Carefully squeeze the area around the puncture point to draw a further blood drop Take care NOT to smear out the blood sample 4 Apply the sample to the test strip j Carefully allow the test strip to come into contact with the blood drop ...

Page 8: ...se contact your local customer support or distributor for assistance 1 Installing the software The software can be downloaded from the website www diatesse nl 2 Connecting the meter to a computer using a USB cable or optionally wirelessly using Bluetooth Windows 10 and up USB cable Insert the cable in an unused USB port on your computer Insert the other end of the cable in the meter s data port wh...

Page 9: ...g The meter will warn you in one of the following ways when the power is low 1 The symbol appears together with notifications on the display the meter is working and the results are still accurate but the batteries must be replaced 2 The symbol appears together with E b Error and LOW There is not enough power to perform a test The batteries must be replaced immediately Replacing the batteries The ...

Page 10: ...d between 10 and 85 relative humidity non condensing for β ketone test strips DO NOT freeze Keep the tests strips in their original tube Do not place in a different tube For foil wrapped test strips do not unwrap until required Store the packaging containing the test strips in a cool and dry place Keep away from direct sunlight and heat Reseal the tube immediately after removing a blood sugar test...

Page 11: ...0 6 mmol L 1 The βOHB concentration can increase if you are fasting and or dieting exercising intensively or suffer from diabetes and fall ill If your β ketone test results in a Lo repeat the test using a new test strip If you receive the same result or if the result does not match how you are feeling then contact your healthcare professional Always follow the recommendations of your healthcare pr...

Page 12: ...ters Check that the code chip has been correctly inserted Check that the code chip used supports the parameters of your meter MESSAGE WHAT IT MEANS WHATTO DO Batteries are dead Insufficient blood sample Test performed incorrectly The control solution bottle was not shaken properly The control solution has expired or is contaminated The control solution is too hot or cold Defective test strip Meter...

Page 13: ...ntdown Automatic switch off after 3 minutes of inactivity Temperature warning Operating conditions Blood sugar 8 C to 45 C 46 4 F to 113 F and 10 to 90 relative humidity non condensing β ketones 10 C to 40 C 50 F to 104 F and 10 to 85 relative humidity non condensing Storage transport conditions for the meter 20 C to 60 C 4 F to 140 F between 10 and 93 relative humidity non condensing Storage tran...

Page 14: ...26 27 NOTES ...

Page 15: ... Borkstraße 10 48163 Münster Germany Import Europa HTMedical BV Distributor GD Medical Pharma BV Hastelweg 224 5652 CLEindhoven info gdmedicalpharma nl TheBluetooth wordmarkandlogosareregisteredtrademarks ownedbyBluetoothSIG Inc andanyuseof suchmarksbyHTMedicalBV isunderlicense Othertrademarksandtradenamesarethoseof their respectiveowners ...