When your new Replaceable Sensor arrives, simply turn off the recorder, remove and 

discard the old sensor, plug in the new one and power the recorder back on.  Your 
recorder will continue to record temperature and humidity without interruption.   
 

USER CALIBRATION 
If you have an accurate standard to compare against, the Replaceable Sensor calibration can be 
adjusted at one point.  This will not adjust the span and is not as accurate as replacing the 
Replaceable Sensor with a newly calibrated one. 

 
1.

 

To activate Calibration mode, turn the unit on and press and hold both the On/Off and the 
Adjust-Down button until the LED is solid green. The LED will then blink amber and at which 
point only the pen being adjusted will

 

show on the display.  

Note:

 

Make sure to press the Adjust-Down button first so the unit will not turn off 

 

2.

 

To raise the unit of measurement, press the Adjust-Down button. To lower the unit of 
measurement, press the Adjust-Up button.  

 

3.

 

When Calibration is complete, press the On/Off button. The adjustment is stored in memory 
even after you turn the unit off or if AC power fails. 

 

Note: After two hours, if no buttons are pressed, the unit will time out of Calibration mode 
and resume normal operation. If you wish to cancel the Calibration, simply enter Calibration 
mode and toggle through the steps without adjusting displayed readings. Exit by pressing 
the On/Off button. You have now restored factory calibration settings. 

 

Note: It is recommended that you use a controlled chamber when determining if Calibration 
is necessary.  Adjusting the pen in an open room is not recommended as air flow can vary 

greatly within a very small area. 

 
 

 
TROUBLE SHOOTING 
Why isn't the chart keeping time or running slow?  

 

The chart may be hung up or restricted, possibly caused by a rip on the outer edge of the 

chart or the chart hub, or the chart may be caught between the arm and pen(s) arm 
platform.  

 

Incorrect chart installed for the selected chart speed.  

 Why did the chart stop turning?  

 

Chart hung up or restricted, (ripped chart)  

 

Unit may be "locked up", this can be confirmed by pressing any of the buttons on the 
keypad, if the unit is locked up there will be no response to button presses and the Chart 
Recorder may appear to be working, but the readings won't change, also the chart will 

not rotate. Remove power and battery, and then re-power.  

 Why don't the display and chart match?  

 

Dip switches set for a specific range, but using chart for another range, or vice versa.  

 

Pen(s) not inserted on pen arm(s) all the way  

 

To adjust pen(s) to match chart see “Pen Home Adjust” in the “Need to Know” section of 

this manual. 

www.

.com

information@itm.com

1.800.561.8187

Summary of Contents for KT6P1

Page 1: ...he timing arrow just to the right of the pen tip on the dial Press Pen Home to move the pen s back onto the chart The following feature should be used for fine adjustments only b To adjust the chart t...

Page 2: ...set the alarm make sure unit is on and press and hold the Alarm button The LED will blink red rapidly for about five seconds then the LED will turn solid green Release the Alarm button and the LED wi...

Page 3: ...ill close on alarm conditions when the alarms are enabled Relay contacts are always functional when the alarm is enabled The relay will close only during minimum and maximum alarm conditions Relays wi...

Page 4: ...d green Release the Pen Home and On Off buttons The LED will flash amber and green for one second then the LED will turn off Note Press Pen Home first so the unit will not turn off 2 The pen will move...

Page 5: ...tion If you wish to cancel the Calibration simply enter Calibration mode and toggle through the steps without adjusting displayed readings Exit by pressing the On Off button You have now restored fact...

Page 6: ...d enough time to recognize this and switch to battery mode The Chart Recorder may lock up or shut off in this situation This situation may also occur if the unit is plugged into an outlet that shares...

Page 7: ...t to operate properly Calibration Only the Replaceable Sensor is calibrated The sensor and all calibration defaults and adjustments are stored on the Replaceable Sensor Accuracy is strictly controlled...

Reviews: