REV 09/2019
Need additional support?
Web:
DicksonData.com
Telephone:
630.543.3747
Email:
support@dicksondata.com
930 S. Westwood Avenue
Addison, IL 60101
KT8
8” TEMPERATURE CHART RECORDER
USER MANUAL &
PRODUCT GUIDE
BATTERY COMPARTMENT
DIP SWITCHES
ADJUST BUTTONS
AC ADAPTER
REPLACEABLE
SENSOR™ PORT
REFERENCE GUIDE
REPLACEABLE SENSOR™
CERTIFICATE OF VALIDATION
Dickson ensures that the Replaceable Sensors™ /
Instruments listed below were developed, tested
and validated together.
Replaceable Sensor™ Model: R400
Instrument Model: KT6P1, KT6P2, KT6P5
Operation:
The Instruments and Replaceable
Sensor™ models listed above were specifically
designed to work together. No other sensors
will work with the Instrument models listed. No
other Instrument will work with the Replaceable
Sensors™ listed. The Replaceable Sensor™ must be
connected to the Instrument for the Instrument to
operate properly.
Calibration:
Only the Replaceable Sensor™ is
calibrated. The sensor, and all calibration defaults
and adjustments are stored on the Replaceable
Sensor™. Accuracy is strictly controlled by
the sensor – no adjustments are made by the
Instrument. Readings are sent from the Replaceable
Sensor™ to the Instrument for storage purposes
only. Certificates of Calibration / NISTs are only
supplied for the Replaceable Sensor™ for this
reason.
Recalibration:
When a Replaceable Sensor™ is due
for recalibration the existing Replaceable Sensor™
can be replaced with a new Replaceable Sensor™
thus eliminating the need to return the entire
instrument for recalibration. Replaceable Sensors™
can be returned for recalibration as well.