Dickson KT8P Manual V6 52418 

NEED TO KNOW

NEED TO KNOW

NEED TO KNOW

NEED TO KNOW    
The pen movement for the KT8P0 is slower than on the AC powered models in order to conserve 
battery power.  In 24-hour mode the pen(s) will move every 1 minute and 30 seconds.  In 7-day 
mode the pen(s) will move every 8 minutes and 30 seconds. 
 
The recorder takes temperature readings every 50 seconds.  An average of the reading taken 
between pen(s) movements is what is drawn on the chart. 
    
TROUBLE SHOOTING

TROUBLE SHOOTING

TROUBLE SHOOTING

TROUBLE SHOOTING    
Why isn't the chart keeping time or running slow?

Why isn't the chart keeping time or running slow?

Why isn't the chart keeping time or running slow?

Why isn't the chart keeping time or running slow?  

 

The chart may be hung up or restricted, possibly caused by a rip on the outer edge of the 
chart or the chart hub, or the chart may be caught between the arm and pen(s) arm 
platform.  

 

Incorrect chart installed for the selected chart speed.  

 Why did the chart stop tu

Why did the chart stop tu

Why did the chart stop tu

Why did the chart stop turning?

rning?

rning?

rning?  

 

Chart hung up or restricted, (ripped chart)  

 

Unit may be "locked up", this can be confirmed by pressing any of the buttons on the 
keypad, if the unit is locked up there will be no response to button presses and the Chart 
Recorder may appear to be working, but the readings won't change, also the chart will 
not rotate. Remove power and battery, and then re-power.  

 Why don't the display and chart match?

Why don't the display and chart match?

Why don't the display and chart match?

Why don't the display and chart match?  

 

Dip switches set for a specific range, but using chart for another range, or vice versa.  

 

Pen(s) not inserted on pen arm(s) all the way  

 

To adjust pen(s) to match chart see “Pen Home Adjust” in the “Need to Know” section of 
this manual. 

 Why does the display read PROB?

Why does the display read PROB?

Why does the display read PROB?

Why does the display read PROB?  

 

Check the sensor pod and KTC probe(s).  Make sure it is properly plugged into the unit. 

 

Is the unit locked up? This can be confirmed by pressing any of the buttons on the 
keypad, if the unit is locked up there will be no response to button presses and the unit 
may appear to be working, but the readings won't change, also the chart will not rotate. 
Remove power and battery, and then re-power. 

 Why does calibration seem to be off?

Why does calibration seem to be off?

Why does calibration seem to be off?

Why does calibration seem to be off?  

 

What is the tolerance of the unit it's being compared to?  

 

It's ok if the unit is within the sum of the two tolerances.  

 

Has an outside calibration house attempted calibration? It may not have been adjusted 
properly.  

 

Walk through calibration adjustment, found in the manual 

 Why won't the battery back

Why won't the battery back

Why won't the battery back

Why won't the battery back----up work?

up work?

up work?

up work?  

 

Are there good batteries in the Chart Recorder? 

 

Keep in mind that battery backup varies significantly depending on temperature, pen 
movement and chart rotation speed.  

 

If there was a momentary power outage (Brown out), the unit may not have had enough 
time to recognize this and switch to battery mode. (The Chart Recorder may lock up or 
shut off in this situation) This situation may also occur if the unit is plugged into an outlet 
that shares the circuit with other machinery that has phase motors or compressors that 
cycle periodically. As these other machines cycle, they momentarily have a high current 
draw, therefore drawing from the unit. 

Summary of Contents for KT8P0

Page 1: ...t time manually by inserting a coin into the groove in the chart hub and turning clockwise until the correct hour and day if applicable on the chart is referenced to the timing arrow just to the right of the pen tip on the dial Press Pen Pen Pen Pen Home Home Home Home to move the pen s back onto the chart The following feature should be used for fine adjustments only b To adjust the chart time pr...

Page 2: ... the pen s are located on the outside edge of the chart press the Pen Pen Pen Pen Home Home Home Home to move pen s to recording position If the pen s are located on the chart press the Pen Pen Pen Pen Home Home Home Home key to move the pen s to the outside edge of the chart A A A Alarm larm larm larm Optional Optional Optional Optional Pen 1 Red Pen Pen 2 Blue Pen F Channel Indicator Low Battery...

Page 3: ...en Pen Pen Home Home Home Home button will scroll through the five options until the A A A Alarm larm larm larm button is pressed to exit from alarm adjust Each press of the Pe Pe Pe Pen n n n Home Home Home Home or Alarm Alarm Alarm Alarm button will store the new settings The unit will take one minute to exit Alarm Set Mode once the Alarm Alarm Alarm Alarm button is pressed 4 4 4 4 If the alarm ...

Page 4: ...ng Time 1 down 2 up 1 down 2 down 1 up 2 up o F 3 down o C 3 up Temperature Ranges Temperature Ranges Temperature Ranges Temperature Ranges 4 down 5 down 6 down 4 down 5 down 6 up 4 down 5 up 6 up 4 up 5 down 6 down 4 down 5 up 6 down 4 up 5 up 6 up 4 up 5 down 6 up 4 up 5 up 6 down All the KT8 recorders feature usre selectable recording times Note Remember to install the correct chart to match co...

Page 5: ...n Home Pen Home Pen Home Pen Home and On Off On Off On Off On Off buttons until the LED is solid green Release the Pen Home Pen Home Pen Home Pen Home and On Off On Off On Off On Off buttons The LED will flash amber and green for one second then the LED will turn off Note Note Note Note Press Pen Home Pen Home Pen Home Pen Home first so the unit will not turn off 2 The pen s will move to the outer...

Page 6: ...T856 calibration can be adjusted at one point This will not adjust the span and is not as accurate as replacing the Replaceable Sensor with a newly calibrated one or returning the KT856 for recalibration 1 To activate Calibration mode turn the unit on and press and hold both the On Off On Off On Off On Off and the Adjust Adjust Adjust Adjust Down Down Down Down button until the LED is solid green ...

Page 7: ...her range or vice versa Pen s not inserted on pen arm s all the way To adjust pen s to match chart see Pen Home Adjust in the Need to Know section of this manual Why does the display read PROB Why does the display read PROB Why does the display read PROB Why does the display read PROB Check the sensor pod and KTC probe s Make sure it is properly plugged into the unit Is the unit locked up This can...

Page 8: ...l number serial number and a PO number ready before calling Replaceable Sensor Certificate of Validation Replaceable Sensor Certificate of Validation Replaceable Sensor Certificate of Validation Replaceable Sensor Certificate of Validation Dickson ensures that the Replaceable Sensors Instruments listed below were developed tested and validated together Replaceable Sensor Model R400 Instrument Mode...