TH800/02/03/05

Temperature & Humidity Recorder

www.DicksonData.com • 1-800-323-2448 • Fax 1-800-676-0498

 

Figure 2

TH802/03/05

Power 

LED

Pen 

Home

Pen 

Home

On/Off

On/Off

Alarm

TH800

Power 

LED

On/Off

The On/Off key turns the unit on and off. 

Pen Home (Down Arrow)

If the pens are located on the outside edge of the chart, 

press the 

Pen Home

 to move pen(s) to recording position. 

If the pens are located on the chart, press the 

Pen Home 

key to move the pen(s) to the outside edge of the chart.

Alarm  Optional

Pen 1: = Red Pen
Pen 2: = Blue Pen

1. 

  To set the alarm, make sure unit is on and press and 

hold the 

Alarm

 button. The LED will blink red rapidly 

for about five seconds, then the LED will turn solid 

green.  Release  the 

Alarm

  button  and  the  LED  will 

turn solid red. At this point the display will show “On” 

or “Off”.

  Pressing  either  the 

Adjust-Up

  or 

Adjust-Down 

buttons, located on the back of the unit next to the 

dip switch, will toggle the alarm on or off.

2. 

  Pressing

  PEN  HOME 

will  scroll  to  the  next  alarm 

options. 

 

Alarm “on” or “off” 

 

 

 

 

  

 

Figure 3

Adjust Buttons Dip Switches

Relay Contacts 

(TH805)

AC 

Adapter 

Jack

Battery  

Compartment

Pen 1 alarm minimum
Pen 1 alarm maximum
Pen 2 alarm minimum
Pen 2 alarm maximum

3. 

 

  

In order to set pen alarm minimums and maximums, 

pressing 

Adjust-Up

  will  increase  the  alarm  value, 

and pressing 

Adjust-Dowm

 will decrease the alarm 

value. There is acceleration if the 

Adjust-Up

 button 

is held down. 

Repetitively  pressing  the 

Pen  Home

  button  will 

scroll through the five options until the alarm button 

is pressed to exit from alarm adjust. Each press of 

the 

Pen Home

 or 

Alarm

 button will store the new 

settings. The unit will take one minute to exit 

Alarm 

Set Mode

 once the 

Alarm

 button is pressed. 

4. 

  If the alarm is triggered, the LED will show as solid 

red and the alarm will sound. Press the 

Alarm

 button 

to silence the audible alarm. 

Alarm Relays

The SPST 24V 500mA relay contacts are normally open and 

will close on alarm conditions when the alarms are enabled. 

Relay  contacts  are  always  functional  when  the  alarm  is 

enabled.  The  relay  will  close  only  during  minimum  and 

maximum alarm conditions. Two pen models have 2 sets 

of relay contacts which operate independently. Relays will 

not operate under battery only power conditions.