Power

The 

TH8

  operates  on  AC  power  with  an  optional  4  AA 

battery backup. Depending on chart rotation selected and 

model; battery backup will last

TH802/03/05

24-Hour Chart Rotation = 1 Day Backup
7-Day Chart Rotation = 2 Day Backup
31-Day Chart Rotation = 1 Week Backup

Note:

  Alarm and relays will not operate under battery 

only power.

TH800

At 70°F 

24-Hour Chart Rotation = 1 Month
7-Day Chart Rotation = 6 Months

At 32°F 

24-Hour Chart Rotation = 3 Weeks
7-Day Chart Rotation = 4 Months

Note:

  (Unit should not be placed in an environment 

below 32°F)

LED Indicators

• AC Power with battery backup–Solid green
•  AC Power with low battery or no battery–Blinks red
• Battery only–Blinks green
• Battery only (Low battery)–Solid red

Note:

 

TH800

 operates on battery only and will blink 

one time every minute to conserve battery life.

CALIBRATION

Your instrument was carefully tested and calibrated before 

being shipped from the factory. For greatest accuracy, we 

recommend factory re-calibration every 6–12 months. Call 

customer service at (630) 543-3747. 
If  you  wish  to  do  calibration  yourself,  follow  these 

procedures. 

1. 

   To  activate  Calibration  Mode,  turn  the  unit  on  

and  press  and  hold  both  the 

On/Off

  button  and 

the 

Adjust-Down

 button (making sure to press the 

Adjust-Down

 button first so the unit will not turn 

off).  The  LED  will  blink  in  amber  rapidly  for  about 

five  seconds,  and  turn  solid  green.  Release  the  

On/Off

 button and the 

Adjust-Down

 button. The 

LED will then blink amber at which point only the pen 

being adjusted will show on the display. 

2. 

   To  raise  the  unit  of  measurement,  press  the  

Adjust-Down

  button.  To  lower  the  unit  of 

measurement,  press  the 

Adjust-Up

  button. 

Pen  home

  switches  between  the  red  and  blue 

pens  on  a  two-pen  unit  and  stores  the  current  

adjustment value. 

3. 

   When  calibration  is  complete,  press  the 

On/Off 

button. Calibration is stored in memory even after 

you  turn  the  unit  off.  User  calibration  information 

will not be lost if AC power fails.

Note: 

After two hours, if no buttons are pressed, the 

unit will time out of user calibration and resume normal 

operation.  The  user  calibration  is  stored  separately 

from the factory calibration. If you wish to cancel out 

your user calibration, simply enter 

Calibration Mode

 and 

toggle  through  the  steps  without  adjusting  displayed 

readings. Exit by pressing the 

On/Off

 button. You have 

now restored factory calibration.

Note:

  It  is  recommended  that  you  use  a  controlled 

chamber  when  calibrating  temperature  and  humidity. 

Calibrating  in  an  open  room  is  not  recommended  as 

humidity can vary greatly within a very small area.

NEED TO KNOW

TH800 Pen Movement

The pen movement for the 

TH800

 is slower than on the 

AC powered models in order to conserve battery power.  

In 24-hour mode the pen will move every 1 minute and 30 

seconds. In 7-day mode the pen will move every 8 minutes 

and 30 seconds. 
The recorder takes temperature readings every 50 seconds.  

An average of the readings taken between pen movements 

is what is drawn on the chart.

Pen Adjust

Pen  adjust  may  be  needed  if  the  pen(s)  and  display  do  

not match.

1. 

   While the unit is on, press and hold both the 

Pen 

Home

 button and the 

On/Off

 button (making sure to 

press the 

Pen Home

 button first so the unit will not 

turn off). The led will blink rapidly between amber 

and green for about five seconds. The LED will then 

turn  green.  Release  the 

Pen  Home

  and 

On/Off 

buttons. The LED will flash amber and green for one 

second and the LED will turn off.