DWL-1900XY Instruction Manual
6
5. Place the device to
position 4
(screen downward). Press “MODE” button once
again and wait until countdown reaches “0”. LCD screen automatically switch to
measuring mode once calibration step 4 is completed.
6. Verify your calibration result by comparing the measurement value of current posi-
tion with the device after turning 180° (in reverse position). If displayed measurement
value differs
>
0.001°
, the device has not been calibrated properly (e.g. surface
unevenness due to rough surfaces or dirt, etc.). Please put a device on smooth sur-
faces to re-calibrate again.
Figure 10. Position 4
Calibration tutorial video available on Digipas website, please visit the following link:
https://www.digipas.com/support/video.php