Precaution:
To achieve maximum accuracy, the device must be held firmly on to jig during calibration process. Any
movement on countdown during each calibration procedure would affect device accuracy.
Position 1
4. Turn the device 180° and place on the opposite side of the
master square as shown in
Position 4
. Hold the device
firmly and do not move the device during each count
down period. Press the MODE button again to start the
calibration and wait until the countdown reach “0”, the
LCD screen displays “Calibration 5” .
5. Rotate the device 180° as shown
Position 5
. Hold the
device firmly and do not move the device during each
count down period. Press the MODE button again to start
the calibration and wait until the countdown reach “0”,
the LCD screen displays “Calibration 6” .
6. Turn the device 180° and place on the opposite side of the
master square as shown in
Position 6
. Hold the device
firmly and do not move the device during each count down
period. Press the MODE button again to start the
calibration and wait until the countdown reach “0”, the
LCD screen displays “Calibration 7” .
7. Place the device to
Position 7
. Press the MODE button
once and wait until the countdown reach “0”, the LCD
screen displays “Calibration 8” .
Position 4
8. Turn the device 180
°
to
Position 8
. Press the MODE button
again and wait until the countdown reach “0”, the LCD
screen will switch to measuring mode once calibration is
completed.
Position 5
Position 8
Position 7
Position 6