Setup Mode
(continued)
Offset Function
The primary display shows the
temperature plus the offset and
the secondary display shows
the offset.
1. When the thermometer is in
Setup mode, the display
shows the
SETUP
icon.
2. Press
/Hold
button or
Max/Min/
button to scroll
to the OFFSET setup option
page (see Fig. 1).
3. Press Enter button to indicate
the offset in the secondary display.
4. Press
/Hold
button or
Max/Min/
button until the
value you want to use appears on the display.
5. Press
Enter
button to store the new setting
into memory.
Using the Offset to Adjust for Probe Errors
Use the offset option in the Setup mode to adjust the
thermometer’s readings to compensate for the errors of
a specific thermocouple. The allowable adjustment
range is ±9.0
°
F or ±5.0
°
C.
1. Plug the thermocouple into the input connector.
2. Place the thermocouple in a known, stable temperature
environment (such as an ice bath).
Fig. 1