23
LOGGING READINGS
The Logging function is controlled by the LOG key.
Prior to logging readings, it is necessary to set up
the time interval between readings (see LOG
SETUP) in the COMPLETE SETUP PROCEDURE
section. The time between readings can be set to
any value from 1 second to 60 minutes. Logging is
initiated by pressing the LOG key. Logging will
continue at the programmed rate until either the
LOG key is pressed again or 1000 logs have
occurred. The display will then show “FULL”.
The STORE function can be used during logging
to insert additional measurements by pressing the
STORE key.
PRINTING
Printing can output real-time readings or stored
readings. The output is sent to the infrared (IR)
printer output located at the top of the thermometer.
The printer output default time period is once every
3 seconds. This time can be set to any value
between 3 seconds and 60 minutes (see PRINT
SETUP) in the COMPLETE SETUP PROCEDURE
section. The following sample printouts show the
RS-232-C format and the HEWLETT-PACKARD
infrared printer in normal and log dump print modes.
Note that in log dump print mode the first line
specifies the total number of readings.
The print function is controlled by the PRINT key.