13
Channel Settings (Continued)
Channel Enable/Disable:
Channel Specific Enable/Disable
options allow the user to turn the
channel on/off.
This provides the user the flexibility
to adjust the settings for as many
channels as they need, but only
measure and log data from ena-
bled ones.
Start/Stop Logging
Here the user can log data points that save internally.
The new data will ALWAYS overwrite
the previous data log!!!
While the data is being logged the unit
must stay on.
Pressing the “Enter” button while on this
screen, while data is being logged, will
stop the data log.
The user should ALWAYS stop the
data log before copying to the USB.
Do not allow the device to run out of
internal memory. There is a table at
the end of the user manual that allows you to see the max time you can log.
View Current Slopes/Offsets:
This function allows the user to view the
slopes and offsets for each of the
calibration points.
A green arrow on the right queues the
user to view the rest of the slopes and
offsets.
(The Up and Down arrows toggle through the
calibration slopes and offset pages.)