G. INVASIVE PRESSURE ALARM LIMITS
• The LifeWindow™ Lite LW8 Invasive Pressure alarm limits default and range is:
Systolic Upper Limit: 160 mmHg Range: 0 to 300 mmHg
Systolic Lower Limit: 60 mmHg Range: 0 to 300 mmHg
Diastolic Upper Limit: 100 mmHg Range: 0 to 300 mmHg
Diastolic Lower Limit: 60 mmHg Range: 0 to 300 mmHg
Mean Upper Limit: 120 mmHg Range: 0 to 300 mmHg
Mean Lower Limit: 60 mmHg Range: 0 to 300 mmHg
Static Upper Limit: 120 mmHg Range: 0 to 300 mmHg
Static Lower Limit: 60 mmHg Range: 0 to 300 mmHg
• In the IP# menu, select the Invasive Pressure Systolic, Diastolic and Mean Upper and Lower
alarm limits. To return to the factory DEFAULT alarm limits, go to the SYSTEM menu and
press the Load Factory Default Settings.
H. INVASIVE PRESSURE ALARM MESSAGES
IP# NO TRANSDUCER – Indicates the Invasive Pressure transducer is not connected to
the unit or a transducer failure.
IP# ZERO TRANSDUCER – Indicates the need to perform a transducer Zero Calibration
procedure.
IP# ERROR ZEROING – Indicates error in transducer Zero Calibration procedure.
IP# LOW SYSTOLIC – Indicates the Systolic value is lower than the selected
Systolic Lower Limit.
IP# HIGH SYSTOLIC – Indicates that Systolic value is higher than the selected
Systolic Upper Limit.
IP# LOW DIASTOLIC – Indicates the Diastolic value is lower than the selected
Diastolic Lower Limit.
IP# HIGH DIASTOLIC – Indicates the Diastolic value is higher than the selected
Diastolic Upper Limit.
IP# LOW MEAN – Indicates that Mean value is lower than the selected Mean
Lower Limit.
IP# HIGH MEAN – Indicates that Mean value is higher than the selected Mean
Upper Limit.
IP# LOW STATIC – Indicates that Mean value is lower than the selected Mean
Lower Limit.
IP# HIGH STATIC – Indicates that Mean value is higher than the selected Mean
Upper Limit.
SECTION 10 – FRACTION OF INSPIRED OXIGEN - FIO2
A. PRINCIPLE OF OPERATION
• The LifeWindow™ Lite LW8 provides fast and accurate oxygen monitoring. It is designed to
monitor up to 100% oxygen concentration in medical gas mixtures. It is capable of rapidly
measuring the oxygen content of an atmosphere or environment.
• The LifeWindow™ Lite LW8 is intended to continuously measure and display the
Concentration of oxygen in a gas mixtures used in medical applications such as Anesthesia
and Respiratory therapy in veterinary patients.
B. PATIENT CONNECTIONS
• The LifeWindow™ Lite LW8 uses one polarographyc oxygen sensor. Attached to the sensor
is a removable plastic diverter. This diverter is used to facilitate the transport of gas mixtures