Digicare Biomedical Technology
Operators Manual Multi-parameter Patient Monitor " LifeWindow Lite LW8 "
59
Input Bias Current : Any lead configuration
<
200 nAdc maximum
Electrode Offset potential :
±
0.5 V
Tall T wave rejection : Rejects T waves
≤
R wave
D. SpO2 - PULSE OXIMETER
SpO2 Range
: 35 to 100% Adult/Neonatal
SpO2 Accuracy
:
±
2% (70 - 100%),
±
3% (50 - 69%) unspecified (0 - 49%)
SpO2 Resolution : 1%
SpO2 averaging : 8 seconds
Pulse Rate Range
: 30 to 250 BPM
Pulse Rate Accuracy : ± 2%
Pulse Rate Resolution : 1 BPM
Pulse Rate Averaging : 16 pulses
Sensor Types
: Finger, Wrap, Y, Universal
E. NON-INVASIVE BLOOD PRESSURE
Method
: Automatic Oscillometric
Measured Parameters
: Systolic, Diastolic, Mean Arterial Pressure, Pulse.
Scale
: mmHg (Kpascal optional)
Operating Modes
: Manual, Automatic, Stat
Auto Repeat Cycles
: 1 - 99 minutes
Stat Cycle
: 1 - 4 minutes
Systolic Range : Adult
60 - 250 mmHg
: Neonatal
40 - 130 mmHg
Diastolic Range
: Adult
40 - 220 mmHg
: Neonatal
20 - 90 mmHg
Mean Range : Adult 45 - 235 mmHg
: Neonatal 35 - 105 mmHg
Cuff Pressure Range
: Adult
0 - 330 mmHg
: Neonatal
0 - 160 mmHg
Absolute Maximum
Inflation Pressure
: Adult 300 mmHg
: Neonatal
200 mmHg
Measurement Time
: Adult 30 seconds typical
100 seconds maximum
: Neonatal 30 seconds typical
80 seconds maximum
Pressure Display Accuracy : ± 3 mmHg
Pressure Resolution : 1mmHg
BP Pulse Rate Accuracy
: ± 3 BPM
40 - 200 BPM
: ± 5 BPM to 255 BPM
Cuffs
: #1, #2, #3, #4, #5, Medium, Large, Extra Large.
: Air Hose with Lure Lock Fitting
F. TEMPERATURE
Range
: 50 - 130
0
F (10 – 50
0
C)
Probe
: YSI
- 400 Compatible, Skin or Rectal / esophageal and
disposable
Scale
: Degrees C (Fahrenheit optional)
Accuracy
:
±
0.1
°
F
from 77 to 103
°
F
(:
±
0.1
0
C from 25 – 45
0
C)
Resolution
: 0.1
°
F
(
0.1
0
C )