Operator's Manual
DS7A - Constant Current Stimulator
Mains Connection
The DS7A is shipped complete with a safety tested mains lead (fitted with a suitable moulded mains plug) to
most countries.
If the mains plug supplied is not suitable please advise Digitimer Ltd of the correct plug for your country.
As it is not possible to obtain certain moulded mains plugs in the UK we will supply a mains lead with a
British plug in these cases.
Mains Plug connections
If you have to remove the fitted plug to connect your own plug, the connections are:-
Line or Hot - Brown
Neutral - Blue
Ground or Earth - Green/Yellow
A mains plug should be used which is so constructed that if the lead is pulled the ground wire cannot come
into contact with the line or neutral pins.
This Instrument MUST be Earthed/Grounded.
In the USA, reliable grounding can only be achieved when the equipment is connected to an equivalent
receptacle marked “Hospital Only” or “Hospital Grade”.
Mains Plug fuse
If it is possible to fit a fuse into the mains plug, this should be of the correct style and be rated at 5 Amps.
Mains Voltage Selector
Check that the voltage selector on the back panel of the instrument is set correctly for your supply voltage.
The factory setting is 230 volts and the alternative is 115V.
The 230V setting allows use on supplies from 200 to 240V
The 115V setting allows use on supplies from 100 to 120V
Digitimer Ltd.
33 of 58
Copyright © 2014
File Ref: N:\Docs\Company\Manuals\DS7A\Issue-14\DS7A-iss14.odt
Last Revised: February 1, 2017