background image

2

 Install the sensor you would like to use into the KF16 port. The following use cases for each sensor   

 are as follows:

         A.) 0.1 Torr CM - The 0.1 Torr Capacitance manometer is used as a standard from 0.1

              Torr to 5 x 10-04 Torr

         B.) The DCP 10 Torr Capacitance Manometer gauge- The DCP has an integral capacitance

               manometer, and is used as a standard in the region of 10 Torr to 0.1 Torr

         C.) 1000 Torr CM- The 1000 Torr Capacitance manometer is used as a standard for the 

               pressure/vacuum region from 1000 Torr to 10 Torr

         D.) The DPP Quantum high vacuum gauge- The DPP employs an advanced pirani 

               technology, and is used for general and high vacuum measurement from 7.5 x 10^(-06)   

               Torr to 1000 Torr for clean, dry environments; and is used as a high vacuum 

               comparison device to validate pressures between  7.5x10^(-06) and 1 x 10^(-03)  Torr.

NOTE : Setting up the DASH

To SELECT your desired sensor (the 0.1 Torr

CM , DCP, 1000 Torr CM, or DPP) power on

the DASH by pressing the PWR/ENT button,

then press the 

 button to open the menu. 

Use  the  ▲ / ▼   buttons  to  toggle  to  Device

Selection, then press PWR/ENT to select. Use

the  ▲ / ▼   buttons  to  toggle  to  the  desired

sensor, then press PWR/ENT to select 

C a l i b r a t i o n

  K I T

1020 Campus Drive, Morganville, NJ 07751 - Office contact: 
732-765-0900 -  sales@digivac.com - www.digivac.com

Quick Start Guide

Connect the 

Bullseye DASH to 

the sensor using 

the appropriate 

cable adapter 

(DB15 for CM's, 

DB9 for Quantum 

Gauges) and 

power on the 

device to monitor 

pressure.

3

4

Reviews: