CLM–70
©
Dinel, s.r.o.
30
Table of default settings
MIN CAPACITY
0 pF
MIN DISPLAY
0%
MAX CAPACITY
300 pF
MAX DISPLAY
100%
UNITS
pF capacity; Display %
DAMPING
2s
FAILURE MODE
LAST VALUE
POLLING ADDRESS (HART
®
)
00
PASSWORD
NO PASSWORD
t
emperatUre
and
pressUre
resistanCe
(high-temperature performance)
performance variant
temperature tm
temperature tp
temperature ta
CLM–70NT–10
-40°C ... +300°C
-40°C ... +200°C
-30°C ... +70°C
CLM –70NT–11(12,13)
-40°C ... +200°C
-40°C ... +200°C
-30°C ... +70°C
CLM–70NT–20(22)
-40°C ... +200°C
-40°C ... +200°C
-30°C ... +70°C
CLM–70NT–30 (31 s KV)
-40°C ... +200°C
-40°C ... +130°C
-30°C ... +70°C
CLM–70NT–31 s PR
-40°C ... +130°C
-40°C ... +130°C
-30°C ... +70°C
CLM–70NT–32
-40°C ... +130°C
-40°C ... +130°C
-30°C ... +70°C
CLM–70NT–61
-
-
-
t
p
–
temperature in
process connection place
t
m
–
medium temperature
on the electrode
t
a
–
ambient temperature
(at the head with electronics)
Fig. 19: Illustration of temperature measurement areas
p
ressUre
resistanCe
(high-temperature performance)
performance variant
maximum overpressure for temperature tp
up to 30°C
up to 85°C
up to 130°C
up to 160°C
up to 200°C
CLM–70NT–10(20, 30)
15 MPa
10 MPa
3 MPa
1)
2 MPa
1)
1 MPa
1)
CLM –70NT–11(12,13,22)
4 MPa
2,5 MPa
2 MPa
1,5 MPa
0,3 MPa
CLM–70NT–32
1 MPa
0,5 MPa
0,1 MPa
-
-
CLM–70NT–61(31)
-
-
-
-
-
1). The above-mentioned values do not apply to hot water, aqueous solutions and steam, in such cases the use must be consulted with
the manufacturer.