2
•
Description
2.1 Overview
Eluent from the gradient pump enters the Rheodyne injection valve
port inside the enclosure. From the injection valve, eluent and
sample flow through the guard column (if used), the separator
column, the suppressor or other post-column device, and finally
through the detector cells.
The LC20 may be configured for optional dual-channel operation. It
will then include a second Rheodyne injection valve and component
panel. There is enough tubing in the LC20 Ship Kit (P/N 046300) to
plumb two systems.
The LC20 does not require power. Internal components are operated
by the GP40/IP20 pump.
2.2 Front Panel
The front panel of the enclosure is the door. It contains access to
one or two Rheodyne injection valves.
2.2.1 Rheodyne Injection Valve
The Rheodyne injection valve is mounted on the component
panel in a manner that allows its control knob to extend
through the front door for access. If the LC20 is configured
as a dual-channel module, a second Rheodyne injection valve
will be installed on the left-side component panel and extend
through the adjacent port.
Doc. 034859-02 9/93
2-3
Summary of Contents for LC20
Page 6: ...Contents iv Doc 034859 02 9 93 ...
Page 8: ...LC20 Chromatography Enclosure 1 2 Doc 034859 02 9 93 ...
Page 12: ...LC20 Chromatography Enclosure 2 2 Doc 034859 02 9 93 ...
Page 24: ...LC20 Chromatography Enclosure 2 14 Doc 034859 02 9 93 ...
Page 26: ...LC20 Chromatography Enclosure 3 2 Doc 034859 02 9 93 ...
Page 29: ...Figure 3 7 Drawing Sample 3 Operation and Maintenance Doc 034859 02 9 93 3 5 ...
Page 32: ...LC20 Chromatography Enclosure 3 8 Doc 034859 02 9 93 ...
Page 34: ...LC20 Chromatography Enclosure 4 2 Doc 034859 02 9 93 ...
Page 40: ...LC20 Chromatography Enclosure 4 8 Doc 034859 02 9 93 ...
Page 42: ...LC20 Chromatography Enclosure 5 2 Doc 034859 02 9 93 ...
Page 50: ...LC20 Chromatography Enclosure 5 10 Doc 034859 02 9 93 ...
Page 52: ...LC20 Chromatography Enclosure A 2 Doc 034859 02 9 93 ...
Page 56: ...LC20 Chromatography Enclosure B 2 Doc 034859 02 9 93 ...
Page 65: ...Figure B 6 Electrochemical Cell Plumbing Schematic B Installation Doc 034859 02 9 93 B 11 ...