5
•
Service
5.1 Introduction
This section describes service and repair procedures for the LC20.
These services are required rarely, but some wear of the fittings and
tubing is to be expected during the life of the enclosure. Before
repairing or replacing any part, refer to Chapter 4 to isolate the
source of the problem. For service and maintenance procedures for
the Rheodyne injection valve, refer to the Rheodyne manual
supplied in the LC20 Ship Kit.
When ordering replacement parts, please include the model and
serial number of your LC20. If possible, also provide the part
numbers and, where applicable, the revision number of the items
you are ordering.
5.2 Tube Fitting Installation
The LC20 is plumbed with Dionex 10-32 ferrule fittings. PEEK
tubing is recommended for use throughout the system. However,
ThermoFlare tubing may be installed where pressures are below
7 MPa (1000 psi); this includes waste lines, tubing from an
autosampler to the Rheodyne valve injection port, or tubing from the
separator column to the suppressor. Use 1.6-mm (0.063-in) ID
Teflon tubing for regenerant lines for both the 2-mm and 4-mm
suppressors.
For instructions on how to install ferrule fittings on PEEK or Tefzel
tubing, refer to Installation of Dionex Ferrule Fittings (Document
No. 034213), provided in the LC20 Ship Kit.
Doc. 034859-02 9/93
5-3
Summary of Contents for LC20
Page 6: ...Contents iv Doc 034859 02 9 93 ...
Page 8: ...LC20 Chromatography Enclosure 1 2 Doc 034859 02 9 93 ...
Page 12: ...LC20 Chromatography Enclosure 2 2 Doc 034859 02 9 93 ...
Page 24: ...LC20 Chromatography Enclosure 2 14 Doc 034859 02 9 93 ...
Page 26: ...LC20 Chromatography Enclosure 3 2 Doc 034859 02 9 93 ...
Page 29: ...Figure 3 7 Drawing Sample 3 Operation and Maintenance Doc 034859 02 9 93 3 5 ...
Page 32: ...LC20 Chromatography Enclosure 3 8 Doc 034859 02 9 93 ...
Page 34: ...LC20 Chromatography Enclosure 4 2 Doc 034859 02 9 93 ...
Page 40: ...LC20 Chromatography Enclosure 4 8 Doc 034859 02 9 93 ...
Page 42: ...LC20 Chromatography Enclosure 5 2 Doc 034859 02 9 93 ...
Page 50: ...LC20 Chromatography Enclosure 5 10 Doc 034859 02 9 93 ...
Page 52: ...LC20 Chromatography Enclosure A 2 Doc 034859 02 9 93 ...
Page 56: ...LC20 Chromatography Enclosure B 2 Doc 034859 02 9 93 ...
Page 65: ...Figure B 6 Electrochemical Cell Plumbing Schematic B Installation Doc 034859 02 9 93 B 11 ...