5. Unscrew the leaking cap from the valve body (Figure 2-3).
Inspect the cap O-ring for cracks and replace if necessary
(P/N 038650).
6. Press the disconnected air line (Step 4) against the small vent
hole between ports 1 and 5. Activate the valve switch to apply
air pressure against the slider. The piston will pop out of the
valve body.
7. Inspect the piston O-ring for cracks and replace if necessary
(P/N 038651).
8. Carefully lubricate the piston O-ring with a small amount of
silicone grease.
Use the grease sparingly. Excess grease may contaminate
the eluent.
9. Press the piston back into the valve body.
10. Make sure that the cap O-ring is inside the cap. Screw the cap
back onto the valve body. Tighten it only fingertight.
11. Reconnect the air line to the valve.
5.6 Cleaning the Column Select Valve Slider
A dirty or plugged valve slider causes excessive system
backpressure and may cause poor chromatographic
performance.
1. Turn off the pump and the eluent selector valves.
2. Remove the plugged valve from its mounting clip (see Figure 2-3
3. Disconnect the air line from one cap. If necessary, use the valve
switch to turn off the air through the disconnected line.
LC20 Chromatography Enclosure
5-6
Doc. 034859-02 9/93
Summary of Contents for LC20
Page 6: ...Contents iv Doc 034859 02 9 93 ...
Page 8: ...LC20 Chromatography Enclosure 1 2 Doc 034859 02 9 93 ...
Page 12: ...LC20 Chromatography Enclosure 2 2 Doc 034859 02 9 93 ...
Page 24: ...LC20 Chromatography Enclosure 2 14 Doc 034859 02 9 93 ...
Page 26: ...LC20 Chromatography Enclosure 3 2 Doc 034859 02 9 93 ...
Page 29: ...Figure 3 7 Drawing Sample 3 Operation and Maintenance Doc 034859 02 9 93 3 5 ...
Page 32: ...LC20 Chromatography Enclosure 3 8 Doc 034859 02 9 93 ...
Page 34: ...LC20 Chromatography Enclosure 4 2 Doc 034859 02 9 93 ...
Page 40: ...LC20 Chromatography Enclosure 4 8 Doc 034859 02 9 93 ...
Page 42: ...LC20 Chromatography Enclosure 5 2 Doc 034859 02 9 93 ...
Page 50: ...LC20 Chromatography Enclosure 5 10 Doc 034859 02 9 93 ...
Page 52: ...LC20 Chromatography Enclosure A 2 Doc 034859 02 9 93 ...
Page 56: ...LC20 Chromatography Enclosure B 2 Doc 034859 02 9 93 ...
Page 65: ...Figure B 6 Electrochemical Cell Plumbing Schematic B Installation Doc 034859 02 9 93 B 11 ...