4. Unscrew one cap from the valve body. Be careful not to lose the
cap O-ring.
5. Press the disconnected air line (Step 3) against the small vent
hole between ports 1 and 5. Activate the valve switch to apply
air pressure against the slider. The piston will pop out of the
valve body.
6. Turn off the air supply.
7. Carefully loosen each of the pressure bolts one-eighth turn to
disengage the port faces from the slider. Do not loosen any
further.
8. Remove the other cap. Push on the slider to force the other
piston out of the valve body.
9. Carefully push the slider out of the valve body with a blunt rod.
10. Inspect the slider surface. If there are scratches in the surface,
replace the slider (P/N 038709).
11. Place the slider in a small container of deionized water or
methanol. Sonicate or agitate vigorously for several minutes.
12. Rinse the slider with deionized water. Blow any water out of the
holes and inspect them for blockage. Use a fine piece of wire
(0.025 in) to dislodge any remaining blockage. Be careful not to
scratch the slider surface. If the blockage cannot be removed,
replace the slider.
13. Push the slider back into the valve body. Orient the dimple on
the top of the slider between ports 1 and 2 of the valve body (see
Figure 2-3).
The valve operates properly only if the slider is correctly
oriented in the valve body. The groove around the valve body
indicates the top of the valve body. The small vent hole is
between ports 1 and 5. The ports are numbered clockwise from
the top around the valve body. The slider connects the ports as
follows:
5
•
Service
Doc. 034859-02 9/93
5-7
Summary of Contents for LC20
Page 6: ...Contents iv Doc 034859 02 9 93 ...
Page 8: ...LC20 Chromatography Enclosure 1 2 Doc 034859 02 9 93 ...
Page 12: ...LC20 Chromatography Enclosure 2 2 Doc 034859 02 9 93 ...
Page 24: ...LC20 Chromatography Enclosure 2 14 Doc 034859 02 9 93 ...
Page 26: ...LC20 Chromatography Enclosure 3 2 Doc 034859 02 9 93 ...
Page 29: ...Figure 3 7 Drawing Sample 3 Operation and Maintenance Doc 034859 02 9 93 3 5 ...
Page 32: ...LC20 Chromatography Enclosure 3 8 Doc 034859 02 9 93 ...
Page 34: ...LC20 Chromatography Enclosure 4 2 Doc 034859 02 9 93 ...
Page 40: ...LC20 Chromatography Enclosure 4 8 Doc 034859 02 9 93 ...
Page 42: ...LC20 Chromatography Enclosure 5 2 Doc 034859 02 9 93 ...
Page 50: ...LC20 Chromatography Enclosure 5 10 Doc 034859 02 9 93 ...
Page 52: ...LC20 Chromatography Enclosure A 2 Doc 034859 02 9 93 ...
Page 56: ...LC20 Chromatography Enclosure B 2 Doc 034859 02 9 93 ...
Page 65: ...Figure B 6 Electrochemical Cell Plumbing Schematic B Installation Doc 034859 02 9 93 B 11 ...