Switch in OFF or LOAD position (slider down):
1
→
4 and 2
→
3
5
→
8 and 6
→
7
Switch in ON or INJECT position (slider up):
1
→
2 and 3
→
4
5
→
6 and 7
→
8
14. Inspect the piston O-rings (P/N 038651) and replace them if
necessary.
15. Carefully lubricate the piston O-rings with a small amount of
silicone grease and reassemble the valve (see Section 5.5 for
instructions to correct Column Select Valve air leaks).
16. Tighten each pressure bolt fingertight, then use an open end
wrench to tighten an additional one-eighth turn.
For optimal leak-free operation at 28 MPa (4000 psi), all eight
pressure bolts must be evenly tightened against the slider
(128 in-oz). A special torque wrench kit (P/N 038943) is
available from Dionex.
17. Turn on the liquid flow and check for leaks. Tighten the pressure
bolts further if leaks appear. Use caution as overtightening may
break off the bolt in the valve body.
18. If cleaning the valve slider does not eliminate the backpressure
problem, see the next section for instructions to replace the port
face.
5.7 Replacing the Column Select Valve Port Face
A scratched port face may cause a leak around the pressure bolt.
A plugged port face will cause excessive system backpressure
and may cause poor chromatographic performance.
1. Turn off the liquid flow.
2. Remove the valve from its mounting clip.
LC20 Chromatography Enclosure
5-8
Doc. 034859-02 9/93
Summary of Contents for LC20
Page 6: ...Contents iv Doc 034859 02 9 93 ...
Page 8: ...LC20 Chromatography Enclosure 1 2 Doc 034859 02 9 93 ...
Page 12: ...LC20 Chromatography Enclosure 2 2 Doc 034859 02 9 93 ...
Page 24: ...LC20 Chromatography Enclosure 2 14 Doc 034859 02 9 93 ...
Page 26: ...LC20 Chromatography Enclosure 3 2 Doc 034859 02 9 93 ...
Page 29: ...Figure 3 7 Drawing Sample 3 Operation and Maintenance Doc 034859 02 9 93 3 5 ...
Page 32: ...LC20 Chromatography Enclosure 3 8 Doc 034859 02 9 93 ...
Page 34: ...LC20 Chromatography Enclosure 4 2 Doc 034859 02 9 93 ...
Page 40: ...LC20 Chromatography Enclosure 4 8 Doc 034859 02 9 93 ...
Page 42: ...LC20 Chromatography Enclosure 5 2 Doc 034859 02 9 93 ...
Page 50: ...LC20 Chromatography Enclosure 5 10 Doc 034859 02 9 93 ...
Page 52: ...LC20 Chromatography Enclosure A 2 Doc 034859 02 9 93 ...
Page 56: ...LC20 Chromatography Enclosure B 2 Doc 034859 02 9 93 ...
Page 65: ...Figure B 6 Electrochemical Cell Plumbing Schematic B Installation Doc 034859 02 9 93 B 11 ...