4. Each column is accompanied by a manual containing
specialized installation and start-up instructions. After
consulting your column manual for any special requirements,
install the columns in the LC20:
a. Remove the end plugs from the separator column and
store them in a safe place. You must reinstall the end
plugs in the columns before placing the columns in
storage.
b. Use PEEK tubing with ferrule fittings to connect the inlet
of the guard column to the #3 port of the Rheodyne valve.
c. Connect the outlet of the guard column to the inlet
(bottom) of the separator column.
d. For suppressed conductivity detection, connect the outlet
(top) of the separator column to the
ELUENT INLET
of the
SRS. Otherwise, connect the separator outlet directly to
the detector inlet.
5. Snap the separator columns into the column clips on the
component panel. See Figure 2-2 for an illustration of the
columns mounted in the clips. The guard columns are held in
place only by the connecting tubing.
B.2.7 SRS Installation (Optional)
1 For complete SRS technical details, refer to the SRS
manual.
2. If you are using an external regenerant:
a. Assemble and fill the regenerant reservoir as described
in the instructions shipped with the reservoir.
b. Locate the 75-mm (30-in) tubing (P/N 035727) in the
SRS Ship Kit. Use this tubing to connect the reservoir
to the
REGENERANT INLET
of the SRS.
c. Adjust the regenerant flow rate as instructed in the
SRS manual.
B
•
Installation
Doc. 034859-02 9/93
B-13
Summary of Contents for LC20
Page 6: ...Contents iv Doc 034859 02 9 93 ...
Page 8: ...LC20 Chromatography Enclosure 1 2 Doc 034859 02 9 93 ...
Page 12: ...LC20 Chromatography Enclosure 2 2 Doc 034859 02 9 93 ...
Page 24: ...LC20 Chromatography Enclosure 2 14 Doc 034859 02 9 93 ...
Page 26: ...LC20 Chromatography Enclosure 3 2 Doc 034859 02 9 93 ...
Page 29: ...Figure 3 7 Drawing Sample 3 Operation and Maintenance Doc 034859 02 9 93 3 5 ...
Page 32: ...LC20 Chromatography Enclosure 3 8 Doc 034859 02 9 93 ...
Page 34: ...LC20 Chromatography Enclosure 4 2 Doc 034859 02 9 93 ...
Page 40: ...LC20 Chromatography Enclosure 4 8 Doc 034859 02 9 93 ...
Page 42: ...LC20 Chromatography Enclosure 5 2 Doc 034859 02 9 93 ...
Page 50: ...LC20 Chromatography Enclosure 5 10 Doc 034859 02 9 93 ...
Page 52: ...LC20 Chromatography Enclosure A 2 Doc 034859 02 9 93 ...
Page 56: ...LC20 Chromatography Enclosure B 2 Doc 034859 02 9 93 ...
Page 65: ...Figure B 6 Electrochemical Cell Plumbing Schematic B Installation Doc 034859 02 9 93 B 11 ...