3.1.1 Sample Loading
The Rheodyne injection valve may be filled completely or
partially. These techniques differ in accuracy, precision, and
the amount of sample required. Refer to the Rheodyne valve
Operating Instructions included in the LC20 Ship Kit before
selecting a method for your application.
Before filling the injection valve sample loop, set the
Rheodyne injection valve knob to
LOAD
(see Figure 3-6)
There are four methods for plumbing the Rheodyne valve
sample loop:
When inserting needles into the needle port, use only
0.028-inch OD (22 gauge) x 2-in. long needles with 90
degree point style (square end). Using the incorrect needle
size can damage the injector.
1. Direct injection into the valve: Set the injection valve to
its
LOAD
position. Insert the syringe into the needle port
on the injection valve (see Figure 3-6). Overfill the
sample loop with several sample loop volumes. Excess
sample will flow out through the waste line. Leave the
syringe in the needle port until the injection valve knob is
turned to
INJECT
.
2. Drawing sample through the waste line:. Place the
injection valve into its
LOAD
position. Insert the syringe
into the needle port on the Rheodyne valve, place the
valve waste line (port #6) into the sample container (see
Figure 3-7). Draw sample into the loop through the
injection valve waste line. No sample will come into
contact with the metal needle of the loading syringe.
Remove the sample from the sample container. Leave the
syringe in the needle port until the injection valve knob
has been switched to
INJECT
.
LC20 Chromatography Enclosure
3-4
Doc. 034859-02 9/93
Summary of Contents for LC20
Page 6: ...Contents iv Doc 034859 02 9 93 ...
Page 8: ...LC20 Chromatography Enclosure 1 2 Doc 034859 02 9 93 ...
Page 12: ...LC20 Chromatography Enclosure 2 2 Doc 034859 02 9 93 ...
Page 24: ...LC20 Chromatography Enclosure 2 14 Doc 034859 02 9 93 ...
Page 26: ...LC20 Chromatography Enclosure 3 2 Doc 034859 02 9 93 ...
Page 29: ...Figure 3 7 Drawing Sample 3 Operation and Maintenance Doc 034859 02 9 93 3 5 ...
Page 32: ...LC20 Chromatography Enclosure 3 8 Doc 034859 02 9 93 ...
Page 34: ...LC20 Chromatography Enclosure 4 2 Doc 034859 02 9 93 ...
Page 40: ...LC20 Chromatography Enclosure 4 8 Doc 034859 02 9 93 ...
Page 42: ...LC20 Chromatography Enclosure 5 2 Doc 034859 02 9 93 ...
Page 50: ...LC20 Chromatography Enclosure 5 10 Doc 034859 02 9 93 ...
Page 52: ...LC20 Chromatography Enclosure A 2 Doc 034859 02 9 93 ...
Page 56: ...LC20 Chromatography Enclosure B 2 Doc 034859 02 9 93 ...
Page 65: ...Figure B 6 Electrochemical Cell Plumbing Schematic B Installation Doc 034859 02 9 93 B 11 ...