Clogged column bed supports.
Replace the bed supports as instructed in the column manual.
•
Columns are contaminated.
Clean the columns (see Column Rejuvenation Procedures,
Technical Note 2R, Document No. 032036).
4.3 Inoperative Column Select Valve
•
Air is not reaching the valve.
Check that the air supply is turned on and is supplying 550 to
820 KPa (80 to 120 psi).
•
Air tubing harness is crimped or blocked.
Check the air tubing for blockage. Remove and replace any
blocked tubing. Reposition crimped or pinched tubes.
•
Air leak.
Air leaks are usually audible; isolate and eliminate leaks.
•
Valve is incorrectly plumbed.
Check that the system is correctly plumbed. Re-plumb the
system if necessary.
•
Pressure bolts on the column select valve are too tight.
Follow the procedure in Section 4.1 to loosen the pressure
bolts if the column select valve is plumbed into the system:
4.4 Column Select Valve Air Leaks
•
Air leak.
Air leaks are usually audible and frequently cause excessive
cylinder gas/air consumption. Locate and repair the leak.
4
•
Troubleshooting
Doc. 034859-02 9/93
4-5
Summary of Contents for LC20
Page 6: ...Contents iv Doc 034859 02 9 93 ...
Page 8: ...LC20 Chromatography Enclosure 1 2 Doc 034859 02 9 93 ...
Page 12: ...LC20 Chromatography Enclosure 2 2 Doc 034859 02 9 93 ...
Page 24: ...LC20 Chromatography Enclosure 2 14 Doc 034859 02 9 93 ...
Page 26: ...LC20 Chromatography Enclosure 3 2 Doc 034859 02 9 93 ...
Page 29: ...Figure 3 7 Drawing Sample 3 Operation and Maintenance Doc 034859 02 9 93 3 5 ...
Page 32: ...LC20 Chromatography Enclosure 3 8 Doc 034859 02 9 93 ...
Page 34: ...LC20 Chromatography Enclosure 4 2 Doc 034859 02 9 93 ...
Page 40: ...LC20 Chromatography Enclosure 4 8 Doc 034859 02 9 93 ...
Page 42: ...LC20 Chromatography Enclosure 5 2 Doc 034859 02 9 93 ...
Page 50: ...LC20 Chromatography Enclosure 5 10 Doc 034859 02 9 93 ...
Page 52: ...LC20 Chromatography Enclosure A 2 Doc 034859 02 9 93 ...
Page 56: ...LC20 Chromatography Enclosure B 2 Doc 034859 02 9 93 ...
Page 65: ...Figure B 6 Electrochemical Cell Plumbing Schematic B Installation Doc 034859 02 9 93 B 11 ...