UltiMate
®
3000 Viper™ Capillary Kits for Intelligent LC (LCi) System Solutions
Quick Installation Guides
Page 10
Rev. 02
January 2010
©
2010 Dionex
Figure 10:
Handling of the eluent filter frit
Note:
To exchange the tubing, first remove the frit, and then remove the eluent
tube before removing the retaining guide.
Important:
Always
use eluent tubing with filter frits for drawing eluents from the
eluent bottles. This prevents contamination of the HPLC system.
Note:
Regularly check the frits for permeability. Especially when working with
aqueous solvents, algae and other microorganisms can grow on the frits
causing clogging. Therefore, exchange the solvents at regular intervals.
Rinse the bottles thoroughly with fresh eluent before reusing them. Replace
the frits (P/N 6268.0110), as necessary.
Filter holder
with filter frit
Retaining guide