UltiMate
®
3000 Viper™ Capillary Kits for Intelligent LC (LCi) System Solutions
Quick Installation Guides
January 2010
Rev. 02
Page 15
©
2010 Dionex
2.4 Flow Schematic
The figure below shows the schematic setup of the tandem operation configuration.
Install your Viper capillaries according to the information of Table 4 and the system stack
schematic (
2.3).
Figure 12:
Flow schematic of tandem operation setup
Table 4:
Detailed description of capillary connections of the tandem operation setup
Description of Connections
Capillary Dimensions (ID x L)
PumpLeft outlet – TCC 10-port valve right, port 8
0.18 mm/0.007” x 950 mm
PumpRight outlet – WPS sampler valve, port 5
0.18 mm/0.007” x 450 mm
TCC 10-port valve right, port 7 – Column 1 inlet
RS: 0.13 mm/0.005” x 250 mm
SD: 0.18 mm/0.007” x 250 mm
TCC 10-port valve right, port 5 – Column 2 inlet
RS: 0.13 mm/0.005” x 250 mm
SD: 0.18 mm/0.007” x 250 mm
TCC 10-port valve right, port 4 – TCC 10-port valve right, port 9
RS: 0.13 mm/0.005” x 150 mm
SD: 0.18 mm/0.007” x 150 mm
TCC 10-port valve right, port 3 – Waste
0.25 mm/0.010” x 1,000 mm
(PEEK)
WPS sampler valve, port 4 – TCC 10-port valve right, port 6
RS: 0.13 mm/0.005” x 350 mm
SD: 0.18 mm/0.007” x 350 mm
Column 1 outlet – TCC 10-port valve right, port 10
RS: 0.13 mm/0.005” x 550 mm
SD: 0.18 mm/0.007” x 550 mm
Column 2 outlet – TCC 10-port valve right, port 2
RS: 0.13 mm/0.005” x 550 mm
SD: 0.18 mm/0.007” x 550 mm
TCC 10-port valve right, port 1 – Detector inlet
RS: 0.13 mm/0.005” x 550 mm
SD: 0.18 mm/0.007” x 550 mm