UltiMate
®
3000 Viper™ Capillary Kits for Intelligent LC (LCi) System Solutions
Quick Installation Guides
Page 26
Rev. 02
January 2010
©
2010 Dionex
5.3 Stacking the Modules
Please stack the modules as shown below. The picture is a schematic to illustrate the course of
the flexible Viper capillaries in general. Dionex recommends installing all capillaries inside
the modules as needed.
SRD-3600
Solvent
Rack and Degasser
DGP-3600RS/SD
Dual-Gradient Pump
WPS-3000(T)SL/
(T)RS
inline Split
Loop Autosampler
Combination of two
stackable Dionex
Detectors
TCC-3000RS/SD
Thermostatted Column
Compartment with 1x 2-
position 6-port switching
valve
The recommended distance between the two stacks is approximately
6 - 8 cm (2.4 - 3.2 inches).
Important:
All modules are primed with 2-propanol at the factory. Prior to initial
operation of the system, make sure that all used solvents are miscible
with 2-propanol and equilibrate your system. Otherwise, perform
appropriate intermediate flushing steps.