DCS_Direct_Jet_1024UVHS_Maintenance_Troubleshooting_Guide_1.2.docx
30 |
P a g e
Hardware Troubleshooting Guide
Use the following table to indentify and fix Mechanical and Electrical faults
Symptoms
Check
Replace
The UV light is not lit.
The light will automatically shut off when the
temperature exceeds 104°F (40°C). It will turn
back on automatically when the operating
temperature has returned to an acceptable value.
The filter. If it is dirty, it will reduce air flow and
can cause high temperature leading to automatic
shut down.
The ambient temperature.
The filter
Switch AC main power on.
No lights on the control
panel and no table or
elevator movement when
control panel buttons are
pressed.
Check AC power cord.
Check AC power source (wall plug).
Check AC power fuses.
Replace power cord.
Change AC power
source.
Replace AC power fuses
Replace Printer Power
Supply