18. Hand Control Plus
TiMOTION
**1. Service indicator – flashes red after 11 months’ operation.
Constant light (red) after 12 months’ operation.
The indicator is a reminder to review the maintenance instruction (see Maintenance
Instructions). This should be done at least once every 12 months.
Reset: hold down buttons <2a> + <2b> for at least 30 seconds.
**The indicator will go out around 30 seconds after the hand control is
activated (battery conservation function).
Linak
The service indicator lights up solid yellow after about 12 months of use. The indicator is a
reminder to go through the Maintenance instruction (should be done once every 12 months).
Reset: Simultaneously hold down the buttons <Fast speed up> + <Fast speed down> for
about 10 seconds until the control box makes a beeping sound as an acknowledgment that
the reset has been done.
The service indicator flashes yellow when the actuator has been overloaded or if another
critical fault has occurred.
Reset: Simultaneously hold down the <Low Speed Up> + <Low Speed Down> buttons for
about 5 seconds until the Service indicator goes out.