CLEANING AND DISINFECTING
After each use, clean the parts to remove any drug residue. This will
prevent ineffective nebulization and reduce the risk of infection.
Washable parts
Nebulizer Kit, Mask (PVC), Mouthpiece
Wash these in hot water or water with vinegar. Rinse thoroughly in clean
and hot tap water, tap gently to remove excess water and allow to dry in
a clean environment. It is recommended to replace the kit after 100 to
120 applications or approximately 20 washing cycles for a single patient.
Non-Washable Parts
Compressor, Air Tube (PVC)
First of all, make sure that the power plug is not plugged into the outlet.
Wipe it with a soft, damp cloth, using water or mild detergent.
Air Filter
Do not wash or clean the air filter. Replace the air filter if it gets wet.
Damp air filters can cause clogging.
Disinfecting
Always disinfect the parts after the last treatment of the day is finished.
If parts are heavily stained, replace them with new ones.
Always wash the baffle after each use.
Do not use a brush or needle to clean the baffle.
Make sure you use plenty of water when disinfecting the parts,
Do not wash the baffle with objects other than the appropriate baffle
accessories.
Warning
:
Use of Baffle
22
Washable
P
arts
Adult
Mask
Children
Mask
Mouthpiece
Medication
Cup
Baffle
Kit Cover
Summary of Contents for MDS-1001
Page 1: ...T rk e 1984 MDS 1001...
Page 5: ...Kit Birle im emas la Haznesi Buharla t r c Kafa Kit Kapa A zl k Maske Hava Hortumu veya 03...
Page 13: ...11...
Page 14: ...12...
Page 19: ...Engl sh Years WARRANTY 1984 MDS 1001...
Page 23: ...KIT SET UP DIAGRAM Medication Cup Baffle Kit Cover Mouthpiece Mask Air Tube or 20...
Page 31: ...28...
Page 32: ...29...