44
6. REPLACEMENT OF BATTERIES
A battery symbol is shown on the display when the batteries are almost empty
. As long as the
stimulator is working normally you can continue the treatment. When stimulation feels weaker than
usual or the stimulator turns off, it is time to replace the batteries.
If the stimulator is not used for some time (approximately 3 months), the batteries should be removed
from the stimulator.
The stimulator operates on either two non-rechargeable 1.5 V AA batteries or two rechargeable 1.2 V AA
batteries, recharged in a separate charger.
The endurance of the batteries depends on the amplitude and number of channels activated during a
program being used. For example, if you use one channel at P1 with an amplitude of 30 mA the non-
rechargeable batteries last approximately 11 hours and rechargeable 7 hours.
NOTE!
It is advisable to have extra batteries at hand during childbirth. Non-rechargeable batteries
might be the better choice, since they last longer than rechargeable batteries.
REPLACING THE BATTERIES
1. Turn off the stimulator.
2. Locate the battery compartment on the back of the stimulator.
3. Remove the battery cover by sliding it downwards about 1 cm and then lifting it upwards.
4. Remove the batteries.
5. Insert the new batteries correctly according to the polarity markings (+ and –) inside the battery
compartment, see picture.
6. Replace the battery cover.
7. Dispose of the exhausted batteries in accordance with local and national regulations.
NOTE!
Non-rechargeable batteries may not be charged due to the risk of explosion.
Summary of Contents for Cefar Femina
Page 29: ...30 ...
Page 57: ...58 58 ...
Page 85: ...86 86 ...
Page 113: ...114 114 114 ...
Page 141: ...142 142 142 ...
Page 169: ...170 170 170 ...
Page 197: ...198 198 198 ...
Page 225: ...226 226 226 ...
Page 253: ...254 254 254 ...