GUIDANCE AND MANUFACTURER’S DECLARATION - ELECTROMAGNETIC IMMUNITY
The VenaPro is intended for use in the electromagnetic environment specified below. The customer or the user of the VenaPro should assure that it is used in such an
environment.
Immunity
Test
IEC 60601 Test
Level
Compliance
Level
Electromagnetic Environment Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
±6kV contact
±8kV air
±6kV contact
±8kV air
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative
humidity should be at least 30%.
Electrical Fast
Transient/Burst
IEC 61000-4-4
±2kV for power
supply lines
±1kV for input/
output lines
±2kV for power
supply lines
±1kV for input/
output lines
Mains power quality should be that of a typical commercial or hospital environment.
Surge
IEC 61000-4-5
±1kV differential
mode
±2kV common mode
±1kV differential
mode
±2kV common
mode
Mains power quality should be that of a typical commercial or hospital environment.
Voltage
dips, short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
<5% U
T
(>95% dip in
U
T
) for 0,5 cycle
40% U
T
(60% dip in
U
T
) for 5 cycles
70% U
T
(30% dip in
U
T
) for 25 cycles
<5% U
T
(>95% dip in
U
T
) for 5 sec
<5% U
T
(>95% dip
in U
T
) for 0,5 cycle
40% U
T
(60% dip in
U
T
) for 5 cycles
70% U
T
(30% dip in
U
T
) for 25 cycles
<5% U
T
(>95% dip
in U
T
) for 5 sec
Mains power quality should be that of a typical commercial or hospital environment. If the user of the
VenaPro requires continued operation during power mains interruptions, it is recommended that the
VenaPro be powered from an uninterrupted power supply or a battery.
Power Frequency
(50/60Hz)
Magnetic Fields
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical
commercial or hospital environment.
NOTE: U
T
is the a.c mains voltage prior to application of the test level.