PT 5201
PTV
Error Codes
Operating Manual
Revision 2
10-1
10
Error Codes
10.1
Command Errors [-199, -100]
Error
Number
Error string
[description/explanation/example]
-101
Invalid character
A command or parameter contains an invalid character, e.g. a header containing
an ampersand, SYST:VERS&.
-102
Syntax error
An unrecognized command or data type was encountered, e.g. a string was
received when the generator didn’t accept strings.
-108
Parameter not allowed
More parameters was received than expected for the command, e.g. the *IDN?;
command accepts no parameters, so receiving *IDN? 2; is not allowed
-112
Program mnemonic too long
The header contains more than twelve characters.
-114
Header suffix out of range
The command is invalid because the value of the numeric suffix attached to the
program mnemonic is out of range, e.g. OUTP:BB12? is illegal because only 3
BBs exist
-121
Invalid character in number
An invalid character for the data type was encountered, e.g. an alpha in a decimal
value.
-124
Too many digits
The mantissa of a decimal numeric data element contained more than 255 digits.
10.2
Execution Errors [-299, -200]
Error
Number
Error string
[description/explanation/example]
-200
Execution error
-222
Data out of range
Indicates that a legal program data element was received but could not be
executed because the interpreted values was outside the range as defined by the
device, e.g. the command OUTP:BB1:SCHP 200; is illegal since the ScHPhase
can not exceed 180deg.
Summary of Contents for PT 5201 Compact VariTime
Page 6: ......
Page 20: ......
Page 40: ......
Page 56: ......
Page 63: ...PT 5210 PTV Operating Manual Revision 2 11 5 END OF DOCUMENT ...