3.5  SIDE PANEL DIP SWITCH SETTINGS 

The dip switches are accessed through the opening on the left side of the MPT-250B.  The access opening is covered 
with a plate that can be swung out of the way.  Turn the large captivated screw counter clockwise until it comes loose 

from the side cover.  The label below the cover depicts the switch banks and lists the switch functions.  The detailed 
description of the switch settings is as follows:  

 

 

 

NOTE:  The unit must be turned off, and then back on for dip switch changes to take effect. 

 

Bank SW1 (Toward rear of tester) 

Switch 1:  

 

 

Not Used 

Switch 2:  

 

 

Not Used 

Switch 3:  

 

 

Not Used 

Switch 4:  

 

 

Not Used 

 
Switch 5: Factory Setting DO NOT adjust. 

 
Switch 6:  Down = Normal 
   

Up = Set point measure enabled 

 
Switch 7: Auto Stop 

 

Down=Off 

   

 

 

 

Up=On 

 
 

Bank SW2 (Toward front of tester) 

Switch 1:  

 

 

Not Used 

 
Switch 2:  

lb units  

Down 

(Section 3.5.2) 

Kg units  

Up 

N units   

Up 

   

Switch 3:  

lb units  

Up 

(Section 3.5.2) 

Kg units  

Down 

   

 

N units   

Up 

   

 

Switch 4:  Factory Setting DO NOT adjust. 

Switch 5: Factory Setting DO NOT adjust. 
Switch 6: Factory Setting DO NOT adjust. 
Switch 7: Factory Setting DO NOT adjust. 

 
3.5.1   AUTO STOP 

This feature stops the pull test when the piece under test fails.  When the force applied to the piece drops below 
70% of the displayed reading, the test is completed and the unit stops.  This feature is turned on or off using 

switch 7 of dip switch bank SW1. 
ON=AUTO STOP enabled. 

 

OFF=AUTO STOP disabled (the STOP switch must be used at the completion of a test). 

  

NOTE:  The AUTOSTOP function will not operate if the test range is at or below approximately 1.5lb.  This is 

necessary to allow for the preload that an operator will normally apply when inserting a connector for the 

test.  If the preload is greater than the lower limit, the unit will not start.  In this case depress the reset 

toggle switch next to the display to be able to begin test. 

 

3.5.2   POUND, KILOGRAM, NEWTON 

Dip switch 2 & 3 of dip switch bank SW2 control the type of units that are measured by the tester.  If pounds are 

desired switch 2 is down and switch 3 is up.  If kilograms are desired switch 2 is up and switch 3 is down.  If 
Newton units are desired both switch 2 and 3 are up. 

 