― 52 ―
― 51 ―
5-10-1 Current (mA/
μ
A) measurement
(m , m , m ,
μ
,
μ
,
μ
Max. rated input current 600 mA dc/ac)
1) What to measure
・
m ,
μ
(DC current): DC circuit current
・
m ,
μ
(AC current): AC circuit current
・
m /m ,
μ
/
μ
(DC current component/AC current component)
・
m /m ,
μ
/
μ
(DC/AC superimposed signal current/AC current
component)
・
Hz (Frequency): Measuring current frequency
2) Measuring ranges
mA : 60.00 mA and 600.0 mA
μ
A : 600.0
μ
A and 6000
μ
A
3) Measuring procedure
①
Set the function selector to
or
, then press the
SELECT button to select [m
@>P
/m
@>P
/m
@
[m
+]@IRUP$UDQJHRUVHOHFW>
[
μ
/
μ
@>
/
μ
@
[
μ
+]@IRU$UDQJH
②
Connect the red plug of the test lead to
measuring
terminal and the black one to the COM terminal.
③
Connect the test pins (red and black) in series with the circuit
to measure.
・
m ,
μ
: Connect the black test pin to the lower electric
potential side of the circuit to measure, and the red test pin
to the higher electric potential side in series with the object.
・
m /
μ
、
m /
μ
: Connect the test pins (red and black) in
series with the circuit to measure.
④
Read the display.
5-10-2 Current (A) measurement
(
, , Max. rated input current AC 10 A dc/ac
)
1) What to measure
・
(DC current): DC circuit current
・
(AC current): AC circuit current
・
/ (DC current component / AC current component)
・
/ (DC/AC superimposed signal current / AC current
component)
・
Hz (Frequency): Measuring current frequency
2) Measuring ranges
6.000 A and 10.00 A
3) Measuring procedure
①
Set the function selector to
, and press the SELECT button
to select a display style from [
@>
/
@>
/
@DQG> +]@
②
Connect the red plug of the test lead to
measuring terminal
and the black one to the COM terminal.
③
Connect the test pins (red and black) in series with the circuit to
measure.
・
: Connect the black test pin to the lower electric potential
side of the circuit to measure, and the red test pin to the higher
electric potential side in series with the object.
・
, : Connect the test pins (red and black) in series with the
circuit to measure.
④
Read the display.
Note:
Measuring
Range
Frequency (Hz)
Input sensitivity(Sine wave)
Frequency range
600.0
μ
A
60
μ
A
15.00 Hz
〜
3.000 kHz
6000
μ
A
600
μ
A
60.00 mA
40 mA
600.0 mA
60 mA