― 76 ―
― 75 ―
Max/Min (with capture mode)
Accuracy:
±(
Specified % rdg + 250 dgt) in each function
Sampling time: Approx. 1 ms
Max/Min (with recording mode)
Accuracy: ±(Specified % rdg + 10 dgt) in each function
(for changes > 100 ms in duration in voltage or current measurement)
How to calculate an accuracy
Example) DC voltage measurement (DC mV)
7UXHYDOXH>P9@
Range accuracy: ±(0.06 %rdg+2 dgt) in the 600.0 mV range
0HDVXULQJHUURU>P9@ðUGJGJW
≒
>P9@
&DOFXODWLRQ>P9@>P9@
5HDGLQJ>P9@WR>P9@
* 2 dgt in the 600.0 mV range corresponds to 0.2 mV.
The product specifications and its appearance described in
this manual are subject to change without prior notice for
improvements or other reasons.
< MEMO >