Dol Sensors
Hedelund 4, Glyngøre 7870 Roslev Denmark Tlf. +45 72 17 55 55 Fax +45 72 17 59 59
www.dol-sensors.com
MOUNTING
Step 1
: Loosen the plastic sensor retainer on the radiation shield by
unscrewing two retainer screws.
Step 2
: Slide the sensor completely into the radiation shield.
Make sure that the sensor is inserted as far as possible. See
Step 3
: Screw the retainer to the radiation shield until the sensor is
firmly held in place. Do not over tighten the screws.
For Pipe Mounting
Step 4:
Remove the U-bolt from the mounting fittings.
Step 5
: Secure the radiation shield to a
3/4˝ to 1-1/2˝ iron pipe.
For Surface Mounting
Step 4
: Remove the U-bolt and the screw holding the pipe-retaining fittings.
Step 5
: Mark the outer two holes of the mounting fittings on the surface.
Step 6:
Drill the two holes marked in the previous step into the surface.
Step 7
: Attach the radiation shield to the surface.
Sensor Retainer
Figure 4: Optional two sided sensor retainer
This two-sided sensor retainer can be used for varying
sensor diameters depending on which side of the retainer is
being used.
MAINTENANCE GUIDE
Clean the radiation shield when needed. At least once a year:
Step 1: Check for damage and miscolouring
– replace if necessary.
Step 2: Remove the sensor from the shield.
Step 3: Clean out spider webs, dust and the like.
Step 4: Clean the top of the shield for any droppings from birds.
Step 5:
Refit the sensor in the radiation shield.
TECHNICAL DATA
PRODUCT SURVEY
Figure 5: Dimensions
140245
Radiation Shield
140200
DOL 12 with 1.4 m cable
S140200A
DOL 12 with 5 m cable
140252
DOL 114 with 2 m cable
140253
DOL 114 with 5 m cable
– TECHNICAL USER GUIDE
130 mm
126 mm
220 mm
120 mm
U-bolt
Sensor diameters
10.2 mm - 12.44 mm
(
0.40” - 0.49”)
Sensor diameters
12.7 mm - 17.5 mm
(
0.50” - 0.69”)
Plastic
sensor
retainer
Figure 3
Summary of Contents for 140245
Page 12: ...604167 2014 12 18 ...