EN
Using the refrigerator
DS 301H, DS601H
18
7
Using the refrigerator
7.1
Before first use
Before first use, clean the refrigerator inside and outside with a damp cloth for
hygienic reasons (see also chapter “Cleaning and maintenance” on page 20).
7.2
Saving energy
•
Only open the medication refrigerator as often and for as long as necessary.
•
Avoid unnecessarily low temperature settings.
•
For optimal energy consumption, position the shelves and drawers according to
their position on delivery.
•
On a regularly basis, make sure the door seal still fits properly.
•
Clean dust and dirt from the condenser at regular intervals.
7.3
Switching on
!
•
At the first starting up of the refrigerator or after switching it off for a long time the
display of the digital thermometer (fig.
, page 4) shows the current ambient
temperature.
•
As soon as the inner temperature reaches the operating temperature range
(2 °C – 8 °C) the current temperature is shown.
1. Reset the thermometer to avoid a wrong maximum temperature value.
2. Press the RESET button for 5 seconds.
7.4
Using the refrigerator
!
CAUTION!
Do not store medicines and medicinal products before the operating
temperature range has been reached.
CAUTION!
Always leave some space between the refrigerated goods in the refrig-
erator to ensure efficient and even cooling. If the drawers are packed too
tightly there may be unacceptable temperature values.
DometicDS301-601H_IOM_4445103317_EMEA16_20xx-xx-xx.book Seite 18 Montag, 22. Februar 2021 6:18 18