5
INSTALLATION
• The TCW3000SDD must be set up in a dry well-ventilated place. Avoid direct sunlight or locating it
close to a heat source. The place must be protected against bearing rain and dust.
• Ensure that the unit is steady and perfectly straight and does not touch any of the units next to it.
• A good air circulation especially around the
compressor is essential for trouble-free
operation. Keep the ventilation openings of the
compressor cover always free from any
obstructions. Make sure that a minimum
distance is provided between the TCW3000SDD
and any wall or other device located next to it.
• Two spacers are included with the
TCW3000SDD. These must be attached to the
rear of the unit as shown in the figure, before
positioning the unit against the wall.
• Do not locate the unit below a ceiling fan or
right next to air-conditioning equipment.
• To ensure proper function, the device must be
standing horizontally. Use a water level to
check and to correct if necessary.
• The TCW3000SDD is equipped with bearing
rolls. Once it is in its final position, the feet
must be screwed up to immobilise it and stop it
rolling away. Screw the feet with the provided
key until the rolls do not touch the floor
anymore.
• Make sure that during operation the drainage
plug is closed properly.
CLEANING / DISINFECTING
• Clean or disinfect the unit inside and outside before using it. Use only gentle cleaning agents. Never
use aggressive or caustic cleaning agents, scouring powder, steel wool, abrasive sponges or
chemical solvents. When cleaning, make sure that no fluids of any kind run into the ventilation
housing.
• For disinfecting, we recommend all the surface disinfecting agents commonly used by the customer,
provided they are recommended by the German Association for Hygiene and Microbiology (DGHM) or
other national organisations. For disinfecting small areas, we recommend using a concentrated
alcoholic agent.
TCW 3000
min.
10 cm
min.
3 cm
foot
bearing
roll
spacer
Summary of Contents for TCW 3000 SDD
Page 2: ......
Page 32: ...Dometic S à r l op der Hei 17 L 9809 Hosingen Luxembourg ...