12
Menu 3 – Customer Settings :
Item
Description
Units
Step
Min
Max
Default
1
Oxygen alarm
--
1
0
1
0
2
Calibration
Calibration software sub section
3
Oxygen alarm level
See oxygen display settings
4
Oxygen alarm delay
S
1
0
600
60
5
Start option enable
--
1
0
1
1
6 Economy
enable
--
1 0
1
1
7
Fault relay on stop
--
1
0
1
0 = No
1 = Yes
8
Set Oxygen Analogue FSD
See oxygen display settings
9 PPM
Offset
0.1
10 Modbus
Address
--
1
1
32
1
11
Remote control
--
1
1
3
1 = local
2 = digital IP
3 = remote com
Menu 4 – Service 1 : Setup
Item
Description
Units
Step
Min
Max
Default
1 Oxygen
sensor
type
Ppm/%
-- ppm
% %
2
Production purity
See oxygen display settings
3 Equalising/non
equalising
1
--
1
0
1
0 = Non Eq.
1 = Eq.
4
Oxygen sensor
--
1
0
1
0 = Digital
1 = RS232
5
Service enable
--
1
0
1
0 = off
1 = on
6
Service due at
--
100
0
999900
0
7
Clear fault log
--
1
0
1
0
8
Pressure sensor fitted
--
1
0
1
0 = Not fitted
1 = Fitted
9
Pressure sensor units
Bar/PSI/MPa
--
--
--
Bar
10
Eco Pressure Hi
Bar/PSI/MPa
0.1
4.1
20
10.0 Bar
11
Eco Pressure Lo
Bar/PSI/MPa
0.1
4
19.9
9.5 Bar
12 Economy
period
M
1 1
120
5
13
Economy
over-ride --
1
0 1 0
=
disabled
1 = enabled
14 Economy
over-ride
period
m
1 1
600
20
15
MODBUS status
--
1
0
1
0 = disabled
1 = enabled
Menu 5 – Service 2 : Timers
Menu
Description
Units
Step
Min
Max
Default
1 Hours
run
H 100*
0
999900 0
2 Hours
economy
H 100*
0
999900 0
3
Hours O2 vent
H
100*
0
999900
0
*1 Hour increments for normal operation, 100 hour increments for adjustment