EN
15
3 Servicing
The recommended Service procedures identified in table 3.2 and all other repair and calibration work should be undertaken by a domnick
hunter trained, qualified and approved engineer.
3.1 Cleaning
Clean the equipment with a damp cloth only and avoid excessive moisture around any electrical sockets. If required you may use a
mild detergent, however do not use abrasives or solvents as they may damage the warning labels on the equipment.
3.2 Service Intervals
Description Of Maintenance Required
Typical Recommended Maintenance
Interval
Component Operation
Daily
2000 Ho
urs
(3-m
on
th
)
4000 Ho
urs
(6-m
on
th
)
80000 H
ours
(12-
mo
nth
)
16000 H
ours
(24-
mo
nth
)
Generator
Check status indicators located on the control panel.
System
Check inlet air quality
Generator
Check for air leaks
Generator
Check pressure gauges during purging for excessive
back pressure
Generator
Check condition of electrical supply cables and conduits
Generator
Check cyclic operation
Generator
Replace active exhaust silencers
Recommended Service A
Filtration
Replace dust and control air filters.
Recommended Service B
Generator
Replace/Calibrate oxygen sensor
Recommended Service C
Generator
Replace/Service valves
Recommended Service D
- Check
– Replace
Typical Recommended Maintenance Intervals
Service
8 000 Hours
(12 Months)
16 000 Hours
(24 Months)
24 000 Hours
(36 Months)
32 000 Hours
(48 Months)
40 000 Hours
(60 Months)
A
B
C
D
domnick hunter ltd and its distributors offer a comprehensive range of preventative maintenance programmes, for further details
contact your local domnick hunter distributor or domnick hunter ltd direct (tel. 44 (0)191 402 9000) for further details.
Summary of Contents for Maxigas 104
Page 1: ...104 120 SERVICE GUIDE ...
Page 2: ......
Page 4: ......
Page 8: ......
Page 10: ...10 Menu Map ...
Page 11: ...11 ...
Page 14: ......
Page 30: ......
Page 32: ......
Page 33: ...EN 33 APPENDIX A SCHEMATICS ...
Page 34: ......
Page 35: ......