• The
Time Active
sets the duration of each series in hh:mm:ss:zzz format. The
Time series
can be used in
combination with a sequence such as the Square Sequence(s) or the Complex Sequence(s) Mode. If the
Time
Active
duration is shorter than the sequence time length, the sequence will stop at the end of the
Time Active
time length.
• The
Interval between series
sets the duration between each series in hh:mm:ss:zzz format.
• The
Total Duration
displays the total duration of the sequence in d:hh:mm:ss:zzz format.
• The
Progression
bar displays the progression of the sequence in %, while the
Time Elapsed
counter displays
the progression in hh:mm:ss:zzz format.
• The
Launch
button starts the sequence.
4.2.2 Configuration Tab
Figure 4.13:
Configuration Tab
The different buttons of the
Configuration Tab
are shown in Figure
and their functions are explained below.
1. The
Add Channel
button opens the
Channels Configuration
window to setup the channels. This window is
detailed in section
2. The
Clear Configuration
button resets the acquisition view and all other parameters set. Any configurations
already set will be lost.
3. The
Save Configuration
button is used to save the Light Source configuration in a
.doric
format.
4. The
Load Configuration
button allows a Light Source configuration in
.doric
format to be loaded. Recorded data
files also contains the configuration used during the experiment and this configuration can be loaded using this
button.
4.2.3 View Tab
Figure 4.14:
View Tab
Chapter 4. Doric Neuroscience Studio
22