DS3
DS3
code: 00
code: 00
MODE
MODE
OK
OK
ON
OFF
ON
OFF
Storing of Moisture Values:
If the storage is activated, by pushing the OK-button up to 250 values can
be stored. If the storage parameter is set to “on single” always one value is
stored, if the parameter is set to “on auto” every value > 0 is stored
automatically until the storage is full or the OK button is pressed again. The
number of measurements per second (mps) can be adjusted between 1 and
10. The storage procedure will be signed with an “stor” on the botom line at
the right side.
Restorage of the Stored Values:
By pushing the
"OK"
and the
“arrow down”
button, the stored values can
be restored. The result is shown in 2 colums. On the left average
value(AV), Minimum value (MI) and maximum value(MA) and on the right
side the stored values, they can be scrolled with the arrow buttons.
With our optional PC-interface cable and software
DMI-Controller
the
stored values can be transfered to a PC for storing them into a file or
showing them on a graph.
DS3
MAT
OK
ON
OFF
SPEICHER:
MA: 8,4
AV: 6,3
MI: 6,1
01: 7,8
02: 7,0
03: 6,1
04: 8,4
DS3
MAT
OK
ON
OFF
6.4%
6.4%
F=
Holzgruppe: 4
HOLD
DS3
MODE
MODE
OK
OK
ON
OFF
ON
OFF
DOSER MESSTECHNIK
87629 Füssen
Tel: 08362 9159402
Fax: 08362 9159407
Cancellation of the Stored Values:
After changing the material adjustment or with code 14, the stored values
will be cancelled, which is shown as “delete storage” on the display.
page 3
Moisture meter type DS5-AL
Parameter Settings:
With code numbers parameter settings can be changed. Press both arrow
buttons, the display shows “code:00“, adjust the required code with the
arrow buttons and confirm it with OK. The following parameters are
changeable:
code 11:
storage adjustments (OFF, ON single, ON auto)
ON single:
single values will be stored by pressing the
OK-button
ON auto:
All values > 0 will be stored automatically after
pressing OK until the storage is filled up or until the OK-button
is pressed again.
code 12:
number of measurements per second (mps)
range 0 - 10
code 13:
number of measurements until automatic switching off (loops)
range 0 - 999, loops = 0: no automatic swiching off
code 14:
delete storage, through changing the material setting the
storage will be deleted too.
code 15:
select language
code 21:
attenuation ( brake), range 0 - 99
code 22:
automatic maximum value measuring (automax)
no automatic maximum measurement (float)
code 95:
toggle between material moisture and calibration
measurements (display of input values)
code 97:
starts the calibration procedure
Turn Off the Meter:
- push "ON/OFF" button until the display shows the service address, this
keeps in the display as long as the “ON/OFF”-button is pressed
- release "ON/OFF" button: the meter turns off
If no measurement is carried out for an adjustable time,
the meter turns off automatically.