background image

DS3

DS3

code: 00

code: 00

MODE

MODE

OK

OK

ON

OFF

ON

OFF

Storing of Moisture Values:

If the storage is activated, by pushing the OK-button up to 250 values can 
be stored. If the storage parameter is set to “on single” always one value is 
stored, if the parameter is set to “on auto” every value > 0 is stored 
automatically until the storage is full or the OK button is pressed again. The 
number of measurements per second (mps) can be adjusted between 1 and 
10. The storage procedure will be signed with an “stor” on the botom line at 
the right side. 

Restorage of the Stored Values:

By pushing the 

"OK"

 and the 

“arrow down”

 button, the stored values can 

be restored. The result is shown in 2 colums.  On the left average 
value(AV), Minimum value (MI) and maximum value(MA) and on the right 
side the stored values, they can be scrolled with the arrow buttons.

With our optional PC-interface cable and software 

DMI-Controller

 the 

stored values can be transfered to a PC for storing them into a file or 
showing them on a graph.

DS3

MAT

OK

ON

OFF

SPEICHER:
MA: 8,4
AV: 6,3
MI: 6,1

01: 7,8
02: 7,0
03: 6,1
04: 8,4

DS3

MAT

OK

ON

OFF

6.4%

6.4%

F=

Holzgruppe:  4

HOLD

DS3

MODE

MODE

OK

OK

ON

OFF

ON

OFF

DOSER MESSTECHNIK
87629 Füssen
Tel:  08362 9159402
Fax: 08362 9159407

Cancellation of the Stored Values:

After changing the material adjustment or with code 14,  the stored values 
will be cancelled, which is shown as “delete storage” on the display.

page 3

Moisture meter type DS5-AL

Parameter Settings:

With code numbers parameter settings can be changed. Press both arrow 
buttons, the display shows “code:00“, adjust the required code with the 
arrow buttons and confirm it with OK. The following parameters are 
changeable:

code 11: 

storage adjustments (OFF, ON single, ON auto)

ON single:

 single values will be stored by pressing the 

OK-button

ON auto:

 All values > 0 will be stored automatically after 

pressing OK until the storage is filled up or until the OK-button 
is pressed again.

code 12:

 

number of measurements per second (mps)
range 0 - 10

code 13:

number of measurements until automatic switching off (loops)
range 0 - 999, loops = 0: no automatic swiching off

code 14:

 

delete storage, through changing the material setting the 
storage will be deleted too. 

code 15:

select language

code 21:

 

attenuation ( brake), range 0 - 99

code 22:

automatic maximum value measuring (automax)
no automatic maximum measurement (float)

code 95:

toggle between material moisture and calibration 
measurements (display of input values)

code 97:

starts the calibration procedure

Turn Off the Meter:

 - push "ON/OFF" button until the display shows the service address, this 

keeps in the display as long as the “ON/OFF”-button is pressed

 - release "ON/OFF" button: the meter turns off
If no measurement is carried out for an adjustable time, 
the meter turns off automatically.

Reviews: