Measurement Preparation: 

 

.

 

 

material thicknesses<5cm 

use suitable bases 

 (for example: polystyrene or foamed plastic plates 

-no metal!) or even better: hold the material to be measured into the air   
.

 

 

thin materials (< 2 cm) 

measure on a pile (thickness at least 2 cm, avoid air gaps between the 

individual layers)   
.

 

 

look for an 

even, smooth surface 

ƒ (minimum size for the measurement 10x10cm)   

.

 

 

minimum distance of the electrodes 

 

 

from the edge of the surface: 1 cm 

 

Turn on the Meter: 

 

.

 

 

Push ON/OFF-button, the display shows the adjusted material group. Hold the instrument into the 

air for automatic zero point measurement and correction, if the zero point is not in the valid range, the display shows 
“NP-ERROR”.  
.

 

 

release the ON/OFF-button, the dislpay shows e.g. "A1: 0,00%" , the instrument is now ready for 

use (the first two characters shows always the adjusted material group and number).   

 

Parameter Settings: 

 

With code numbers parameter settings can be changed. Press both arrow buttons, the display shows “code:00“, 
adjust the required code with the arrow buttons and confirm it with OK. The following parameters are changeable: 
code 11: storage adjustments (OFF, ON single, ON auto)   

ON single: 

single values will be stored by pressing the OK-button 

ON auto: 

All values > 0 will be 

stored automatically after pressing Ok till the storage is filled up or till the OK-button is pressed again.   

code 12: number of measurements per second (mps) range0-10 code 13: number of 

measurements till automatic switching off (loops) range 0 -999, loops = 0: 
no automatic swiching off code 14: delete storage, through changing the 
material setting the   

storage will be deleted too. code 21: attenuation ( brake), range0-99 code 22: automatic maximum 

value measurinhg (automax)   

no automatic maximum measurement (float) 

 

Material Groups: 

 

For using the required stored material group the corresponding code 
has to be selected (the sign is always displayed as the first character): 
code 30: sign S, customer specific curves 
code 31: sign H, wood (HD5) 
code 32: sign P, paper and cardboard (PD2) 
code 33: sign A, wood (BD2) 
code 34: sign B, building materials (BD2) 
 

the selected group is always displayed during the measurements 
 

Additional adjustments can be made with the optional PC-Software DMI-Controller 

Checking Measuring Quality: 

 

Summary of Contents for DM 4 A

Page 1: ...Feuchtigkeitsmessger te Moisture Meter Humidim tre...

Page 2: ...r wood building materials paper and cardboard Description The electronic moisture meter DM4A is used to determine in a matter of seconds the moisture in materials The average moisture down to a depth...

Page 3: ...he meter works in accordance with the principle of an opened plate capacitor The capacity of the capacitor depends on the material dielectric constant of the material in between the plates Compared wi...

Page 4: ......

Page 5: ...ow buttons and confirm it with OK The following parameters are changeable code 11 storage adjustments OFF ON single ON auto ON single single values will be stored by pressing the OK button ON auto All...

Page 6: ......

Page 7: ......

Page 8: ...higher depth The measuring result is an average moisture value With our moisture meter it s possible to observe the drying process If at the same place after several days the moisture result does not...

Page 9: ...nearisation curves you need the PC interface cable and the software DMI Controller The customer specific curves can be selected with code 30 The material group is the spec curves Basic Calibration The...

Page 10: ...or higher moisture content Attention Frames may also causes higher moisture readings Therefore the instrument also can be used to detect frames Moisture measurement in plastic boates Caused through Os...

Page 11: ...Attention Frames or moutings may also causes higher moisture readings Therefore the moisture meter also can be used to detect frames or mountings at the opposite side of the wall...

Page 12: ...checking and calibrating the moisture meter state the type of the meter plastic case for better protection different types of moisture meters special calibration of the moisture meter interface cable...

Page 13: ...roducts and their uses They should not be taken to imply special characteristics or suitability for any specific purpose other than those stated We constantly work to improve our products and reserve...

Reviews: